1.3 Specifications
Power Requirements
Refer to Section 1.2 Model Number Breakdown.
Power Consumption
Refer to Signal Converter Instruction Bulletin.
Flowmeter
Characteristics
Meter Size/Flow Ca-
pacity
Refer to Table 1-4.
Span
Factory set at specified range between extremes listed in
Table 1-4; can be field adjusted.
Rangeability
100:1
Minimum Liquid
Conductivity
5
µ
S/cm
System Accuracy
Refer to Signal Converter Instruction Bulletin.
Meter Capacity
Specified on Flowmeter data tag (equal to maximum flow capac-
ity in engineering units). Refer to Table 1-4.
Environmental Limits
Ambient Temperature
Models 10DX3111:
-13
o
to 140
o
F* (-25
o
to 60
o
C *)
Model 10DX3311:
-13
o
to 122
o
F (-25
o
to 50
o
C)
* For Process Temperatures up to 266
o
F ( 130
o
C) , maximum temperature is re-
duced by one degree for each degree increase in Process Temperature above 266
o
F ( 130
o
C).
Relative Humidity
10% to 90%
Series 10DX3000 Magnetic Flowmeter Instruction Manual
1-8