By using the standard “Copy and paste“ feature, you can now transfer the data
for further processing, for example, into a spreadsheet application.
ATTENTION:
Please note that depending on the size of data, downloading and
displaying all data can take some time. dBm values are transfered with two deci-
mal places.
ATTENTION:
Please note that the internal 64K memory can quickly get filled up.
Consequently, we
strongly
recommend our 1MB memory expansion (
Option
001
) for using the logger-function.
7.22 RunPrg
(Run program)
SPECTRAN features its own file system which is loaded with various programs.
You can even write your own programs and store them right there. Also, saved
SETUPs (see the following paragraph) use this program feature. Every program
is stored with a specific number. When using this function, any of these pro-
grams can be recalled and executed by just typing its number.
7.23 Setup
(Configuration)
Use
Setup
to manage individual configurations and programs.
The
Store
option will save the current configuration (fLow, fHigh, RBW, VBW,
SpTime, Atten, RefLev, Range, Detec) as its own little program. Please enter a
number
above 200
to store configurations. The previously described
RunPrg
function can be used to recall a configuration.
Del
deletes a previously stored
configuration: please just enter the corresponding program number.
Factor
recalls the factory settings.
This option is especially useful for unex-
perienced users having “messed up“ the device’s configuration.
Attention:
Please note that you can only store full kHz frequency values with the
Store
comand but if you use our free LCS PC-software there is not such a limi-
tation. With the PC software you can store much more values like used sensor,
unit, scale etc.
HINT:
When using program numbers
100
to
109
, you can directly recall the cor-
responding program using
keys 0 to 9
! However, the previous key assignment
will be overwritten.
7.0 The main menu
Firmware V 1.0 / © 2005-2013 by Aaronia AG, D-54597 Euscheid, www.aaronia.com
32