AD-4405 Weighing Indicator
Page
17
ESC
key ..... The key to store the parameters temporarily. Proceed to the
Cal 0
display.
Press and hold the
ON/OFF
key and press the
ESC
key
No parameters are changed,
CanCel
is displayed and the calibration
mode is finished.
Step18 Press
the
ON/OFF
key to turn the display off.
5.3.
5.3.
5.3.
5.3. Weighing Range Function
Weighing Range Function
Weighing Range Function
Weighing Range Function
The weighing range function can select "single range" or "dual range". Specify each
weighing interval (division) for the multi-interval instrument. Each weighing interval is
displayed according to a net value or gross value.
Caution
When single range is used, performing this function is not required.
Example 1
The gross display.
Specified
parameters:
First range
Range = 50.00kg, division 0.02kg
Second range
Range = 100.00kg (maximum capacity), division 0.1kg
Display
0kg to 50kg :
The first range, division 0.02kg.
50kg to 100kg :
The second range, division 0.1kg.
Division
=
0.02kg
Division
=
0.1kg
0kg 50kg 100kg Gross
Weighing
value
Example 2
The net display using a 40kg tare value.
Specified parameters:
The same parameters as example 1.
Display
-40kg to 50kg :
The first range, division 0.02kg.
50kg to 60kg :
The second range, division 0.1kg.
Division = 0.02kg
Division=0.1kg
-40kg 0kg 50kg 60kg Net
Tare value
90kg 100kg Gross
40kg Weighing value
5.3. Weighing Range Function
Summary of Contents for Weighing Indicator AD-4405
Page 1: ...Weighing Indicator WM PD4000504...
Page 76: ...Page 74 AD 4405 Weighing Indicator MEMO...
Page 77: ...AD 4405 Weighing Indicator Page 75 MEMO...
Page 78: ...Page 76 AD 4405 Weighing Indicator MEMO...
Page 79: ...13...
Page 80: ......