background image

13

Operating

 

The Bluetooth Function

Communications

 

Interface:

To connect to a new device (new master) the

must

 

be

 

out of range

 

from the previous

 

master

 

or unpaired

 

from that

 

master.

UT-302PBT

Using Bluetooth communications terms, the

UT-302PBT is a slave device. To connect the UT-302PBT 

to a master device you must initiate the connection by 

first pairing with the UT-302PBT. To complete the pairing 

process the “0000” number must be provided to the 

master as the Bluetooth  Pass Key (Bluetooth PIN Code). 

Once the pairing is complete the UT-302PBT will 

automatically reconnect to the master device when 

possible. The four digit Bluetooth Pass Key must be 

entered only during the pairing to a new master

Summary of Contents for UT-302PBT

Page 1: ...PBT DIGITAL BLUETOOTH THEROMETER ...

Page 2: ...of LCD Display Battery Installation Changing Modes C F Switchable Tips for Measuring Human Temperature Measuring Body Temperature in the Ear Memory Function Care and Maintenance Applied Standards Error Codes Technical Specifications FCC Statement EMC Tables Operating The Bluetooth Function 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 ...

Page 3: ... advantages include 1 Wireless communication Bluetooth class 2 2 Intended use For home clinic and hospital use to determine human body temperature 3 Illuminated back light display 4 10 user memory 5 C F switchable function 6 Probe cover free 7 Waterproof probe design 8 One second reading 9 Auto power off for power saving 10 Low battery indicator 11 Warning indication for battery condition and meas...

Page 4: ...ould stay in steady state room condition for at least 30 minutes 9 Avoid measuring temperature in 30 minutes after exercise bathing or returning from outdoor 10 To protect the environment dispose of empty batteries at appropriate collection sites according to national or local regulations 11 Please use the thermometer solely for its intended purpose Carefully hold the device when in use to avoid d...

Page 5: ... Degrees of Temperature Celsius Scale Fahrenheit Scale Scan in Progress Low battery Warning Memory Symbol and Sets of Records Smiling face Indicating temperature is lower than 38 C 100 4 F Frowning face Indicating temperature is equal or higher than 38 C 100 4 F 3 LED Wireless connecting indicator Wireless Reset Key Wireless Power Button ...

Page 6: ...y 1 Use a Phillips head screwdriver to loosen battery cover screw Remove the battery cover 2 Insert or replace 2 x 1 5 V AAA alkaline batteries into the battery compartment ensuring to match the indicated polarity symbols Always use new batteries Battery operated 1 Please properly dispose of the batteries away from small children and heat 2 It is recommended to remove the batteries if the unit wil...

Page 7: ...ermometer can display result in either degrees Fahrenheit F or degrees Celsius C Changing C F please make sure the device is in power on condition Press the Power Scan button hold it and then press the Memory button to toggle between C and F ...

Page 8: ...is often too low The measurement must not be taken in an ear affected by inflammatory diseases e g discharging pus or secretion after possible ear injuries e g eardrum damage or in the healing period after operative procedures In all of these cases please consult your doctor Use of the thermometer on different persons can be inappropriate in the event of certain acute infectious diseases because o...

Page 9: ...d backwards before inserting the sensor tip This is important so that the sensor tip can be pointed directly at the eardrum 4 Insert the probe into the ear canal the probe must insert well to ensure an accurate temperature measurement then press the Power Measure button for about 1 second release the button and you will hear 1 short beep sound means complete the measurement 1 Remove the front cove...

Page 10: ...ut off after one minute without use Measuring Body Temperature in the Ear 1 To ensure accuracy please wait at least 20 seconds between successive readings 2 Clean the thermometer lens again after use Wait at least 10 minutes between consecutive readings to ensure accuracy Under 1 year Have child lay flat with the head sideways so that the ear is facing upwards Gently pull the ear straight back 1 y...

Page 11: ...ing Body Temperature in the Ear 9 The Readings If the temperature measurement is below 38 C a Smiling Face will be appear next to the reading If the reading is 38 C or above a Frowning Face will be displayed ...

Page 12: ... next previously stored measurement so then all the way to 10 Memory Function You can recall up to 10 measurements of all currently stored measurements in memory to share with your physician or trained healthcare professional Memory Recall 3 Automatically on the 11th measurement when the 10 memories have been used up any new measurement will be recorded and the oldest memory deleted ...

Page 13: ... When the device connects successfully to health monitoring devices and Bluetooth capable devices such as a computer PDA or mobile phone the LED will remain lit At conclusion of temperature measurement the reading will be sent to the health monitoring device and the reading will automatically stored in memory 3 To disconnect from health monitoring devices turn off the wireless power From on to off...

Page 14: ...A etc Please refer to the instruction manual of your personal computer for how to install Serial Port Profile SPP Bluetooth LED SLOW BLINKING indicate is waiting for connection from other master Bluetooth device ACTIVE indicate Bluetooth connection have been established successfully FAST BLINKING indicate this UT 302PBT is at linking state waiting be linked by other Bluetooth terminal system like ...

Page 15: ...slave device To connect the UT 302PBT to a master device you must initiate the connection by first pairing with the UT 302PBT To complete the pairing process the 0000 number must be provided to the master as the Bluetooth Pass Key Bluetooth PIN Code Once the pairing is complete the UT 302PBT will automatically reconnect to the master device when possible The four digit Bluetooth Pass Key must be e...

Page 16: ...IEC EN 60601 1 2 Clinical thermometers Part 5 Performance of infrared ear Thermometers with maximum device Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature Medical electrical equipment Part1 General requirements for safety Medical electrical equipment Part2 Collateral standard Electromagnetic compatibility Requirements and tests Applied Standar...

Page 17: ... o o 0 2 C 0 4 F 0 3 C 0 5 F LCD Display Cause Solution The temperature measured is lower than 34 C 93 2 F Operate the thermometer only between the specified temperature ranges If necessary clean the sensor tip In the event of a repeated error message contact your retailer or Customer Services The temperature measured is higher than 43 C 109 4 F The operating temperature is not in the range from 1...

Page 18: ...use harmful interference to radio or television reception which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one of the following measures 1 Reorient or relocate the receiving antenna 2 Increase the separation between the equipment and receiver 3 Connect the equipment into an outlet on a circuit different from that to which the ...

Page 19: ...t accept any interference received including interference that may cause undesired operation 7 This device complies with FCC radiation exposure limits set forth for an uncontrolled environment End users must follow the specific operating instructions for satisfying RF exposure compliance This transmitter must not be co located or operating in conjunction with any other antenna or transmitter ...

Page 20: ...ed for domestic purposes Thermometer Harmonic emissions IEC 61000 3 2 Not applicable Voltage fluctuations flicker emissions IEC 61000 3 3 Not applicable Guidance and manufacturer s declaration electromagnetic immunity The is intended for use in the electromagnetic environment specified below The customer or the user of the should assure that it is used in such an environment Thermometer Thermomete...

Page 21: ...ld be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol Thermometer NOTE 1 NOTE 2 At 80 MHz and 800 MHz the higher frequency range applies These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Field strength...

Page 22: ...istance according to frequency of transmitter m 150 kHz to 80 MHZ d 1 2 P 80 MHz to 800 MHZ d 1 2 P 800 MHz to 2 5 GHz d 1 2 P For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W a...

Page 23: ...This device includes RF transmitter with wireless communication 21 ...

Page 24: ...ess Park Abingdon Oxfordshire OX14 1DY United Kingdom Telephone 44 1235 550420 Fax 44 1235 550485 A D ENGINEERING INC 1756 Automation Parkway San Jose California 95131 USA Telephone 1 408 263 5333 Fax 1 408 263 0119 A D AUSTRALASIA PTY LTD 32 Dew Street Thebarton South Australia 5031 AUSTRALIA Telephone 61 8 352 3033 Fax 61 8 352 7409 2014 A D Medical All rights reserved ...

Reviews: