Page 10
Functions and Specifications
Specifications
Performance specifications
Measurement method
Oscillometric
Pressurization
Display range
0 ~
320 mmHg
Interval measurement
85 ~
300 mmHg (Fitted)
Manual measurement
185 mmHg (Fixed)
Measurement range
Systolic Blood Pressure
60 ~
280 mmHg
Diastolic Blood Pressure
40 ~
160 mmHg
Pulse rate
30 ~
200 bpm
Accuracy
Pressure
±
3 mmHg
Blood pressure Conforming to 1992 AAMI standard
(
±
3 mmHg or
±
2% measurement whichever is greater)
Pulse rate
±
5 %
Minimum display division Pressure
1 mmHg
Pulse rate
1 bpm
Measurement
Automatic measurement
Manual measurement
Number of measurements Apporx. 180 (Actual measurement number may vary
due to environment and capacity of batteries)
Memory
Up to 300 sets of data
Normal
Clock
Display
During a measurement
Pressure value
After a measurement
SYS, DIA and pulse rate
Error code, function of concealing the measurement data
Clock
24-hours (1997~2096 year, automatic leap year setting)
Batteries
3 x Alkaline battery (type LR6, type AA) or
3 x Ni-Cd battery (type AA)
Type of protection
Internally powered equipment type BF
against electric shock
CE Marking
The label of the medical device
by the EC directive.
0366
C-Tick Marking
The certification trade mark registered
to the ACA by the Trademark office.
AAMI : Association for the Advancement of Medical Instrumentation
ACA: the Australian Communications Authority
CE marking and C-Tick marking are labeled only where they are required.
N92