Page 24
Operation
Automatic Measurement
Caution
❏
Automatic measurement uses the internal clock and parameters of automatic
measurement. Refer to section "Parameters for the Display and Clock" and
"Selection for the automatic measurement" for setting these parameters.
❏
Press and hold the
AUTO ON/OFF
key for approx. 3 seconds so as to turn off "
A
" symbol
on display, when the patient stops the automatic measurement or detaches the cuff.
If the automatic measurement mode is running without a patient arm or resistance,
damage will occur to the cuff.
Starting or re-starting automatic measurement mode
Step 1
Confirm the parameters for automatic measurement. Refer to "Selection for
the automatic measurement".
Step 2
Press and hold the
AUTO ON/OFF
key for about 3 seconds. When the "
A
" is
displayed the recorder starts an automatic measurement based on the
internal clock and the parameters for automatic measurement.
Operation using mode
II
(Sleep mode)
Step 1
Press the
AUTO ON/OFF
key turning off the "
S
" when the patient wakes up.
Step 2
Press the
AUTO ON/OFF
key turning on the "
S
" when the patient goes to bed.
Stopping or canceling automatic measurement
Step 1
When the
AUTO ON/OFF
key is pressed for about 3 seconds the recorder exits
the automatic measurement mode and the "
A
" disappears. Conversly, by
pressing the key again the "
A
" will reappear.
Manual Measurement
Step 1
Press the
START STOP
key. The recorder starts a measurement at once. The
results are displayed and stored in memory.
To stop a Current Measurement
Step 1
Press the
START STOP
key during measurement. The recorder will stop the
measurement at once and releases the air from the cuff.
Automatic Measurement
Operation
AUTO ON/OFF