28
13.3. Function list
Item
Parameter
Description
Auto power-off
function
poff
0
Auto power-off disabled
Turns the power OFF
automatically
1
Auto power-off enabled
Weight display
resolution
re5o
(See Note)
0
Normal (1/3,000)
Changes the
minimum display
1
High (1/6,000 or 1/7,500)
2
Maximum (1/12,000 or 1/15,0 00)
Weighing unit
Ut-g
0
kg
Only SW-6K,15K and
30K can be set
1
g
Zero tracking
trc
0
Zero tracking function disabled
Tracks the zero drift
1
Zero tracking function enabled
Stability band
width
5t-b
(See Note)
0
±0.5d (width 1d)
Conditions to turn the
STABLE indicator on
d = minimum display
in kg
1
±1.0d (width 2d)
2
±2.0d (width 4d)
Stability
detection time
5t-t
0
0.5 second
1
1.0 second
2
2.0 seconds
Stability
Response speed
Cond
0
Weak stability / fast response
Filtering. Factory
setting for models of
weighing capacity 60
and 150 kg is
1
.
1
Normal stability / normal response
2
Strong stability / slow response
Serial interface
baud rate
bp5
0
2400 bps
Only when the options,
RS-232C or RS-422/485,
are used.
1
4800 bps
2
9600 bps
Serial interface
data bits / parity
btpr
0
7 bits / EVEN
1
7 bits / ODD
2
8 bits / Non parity
Serial interface
Data output
mode
prt
0
Stream mode / Command mode
1
Command mode only
2
Print key mode / Command mode
3
Auto-print mode +/- data / Command mode
4
Auto-print mode + data/ Command mode
5
Auto-print mode +/- data and OK
/ Command mode
6
Auto-print mode + data and OK
/ Command mode
Serial interface
Type
5if
0
RS-232C
1
RS-422
2
RS-485
Serial interface
Address
adr
##
## = 00 to 99