15.2 Parameter List
Item
Display
Meaning and purpose
?
The type L scale is turned off after 5 minutes of no operation,
when displaying zero.
Automatic
f1 0
#
Not used
?
turning off
f1 1
Used
?
?
Transmission rate of the serial interface (RS-232C/ 422/485).
f4 0
#
2400bps
?
Baud rate
f4 1
4800bps
?
f4 2
9600bps
?
Mode selection for the serial interface (RS-232C/ 422/485).
f5 0
#
Stream mode. (Refer to “16.2 Stream Mode”)
?
f5 1
Command mode. (Refer to “16.3 Command Mode”)
?
f5 2
A data is output, when the
switch is pressed.
?
f5 3
Auto-print +
When the display becomes a positive stable value above
near-zero, the scale outputs the data automatically. Next
Output mode
output can be performed after the display becomes near-
zero or a negative value.
?
?
f5 4`
Auto-print +/-
When the display becomes a stable value outside near-zero,
the scale outputs the data automatically. Next output can be
performed after the display becomes near-zero.
?
?
f5 5
At each finish of full/dribble batch function, the data is output.
?
Mode selection for the comparator
f6 0
#
Pressing the
SET
switch, the scale always compares the
current display value.
A
A
f6 1
The scale always compares the display value.
A
Comparator
f6 2
Pressing the
SET
switch, the scale always compares the
display value when not near-zero.
A
f6 3
The scale always compares the display value when not
near-zero.
A
Notes
The “#” shows factory settings.
The “near-zero” is within
±
4 digits from zero point in the unit of kg.