12
When Pound/Ounce is selected as a weighing unit, enter the values in ounces for
comparison.
In the density mode, comparison is performed to the density obtained.
3-4 Example of Use
The following example uses the AD-8951 comparator light, which is sold separately, to display the
comparison result in red, green, or orange.
1 Using the AX-KO507-W200 cable sold separately, connect the comparator light to the balance
as shown below.
2. Set the “Comparator (
Cp fnc
)” of the balance function table as follows:
Cp
3
Compares the result excluding the data near zero continuously.
Cpin
0
Inputs the upper or lower limit values digitally.
bep
1
Sounds the buzzer for LO.
bep
0
Does not sound the buzzer for OK.
bep
1
Sounds the buzzer for HI.
3. Set the upper and lower limit values as follows:
Cp Hi
101.000 g (Upper limit)
Cp lo
99.000 g (Lower limit)
4. The comparator and buzzer functions as follows, depending on the comparison result.
Weighing data
Comparator light
Buzzer
90.000 g
Orange
Sounds
100.00 g
Green
Does not sound
110.000 g
Red
Sounds
Summary of Contents for GX Series
Page 1: ...WM PD4000216B...
Page 8: ...6...
Page 9: ...7...
Page 24: ...22 4 7 Data Format Examples...
Page 35: ...33...