background image

 

 18 

Storing tolerance range from center value using numerical input 

Step 10 Press the  PRINT  key while displaying  

  . The current parameter is displayed.   

If the tolerance range is to be changed, change it using the following keys. 
Input the tolerance range with center value assumed to be 

100

%.  

 SAMPLE  key 

Moves blinking digit. 

  RE-ZERO    key 

Changes value of blinking digit. 

 PRINT  key 

Stores and proceeds to step 11. 

 CAL  key 

Cancels and proceeds to step 11. 

Step 11  Press the  PRINT  key while displaying  

  . The current parameter is displayed.   

If the second tolerance range is to be changed, change it using the following keys. 
Input the second tolerance range with center value assumed to be 

100

%.  

 SAMPLE  key 

Moves blinking digit. 

  RE-ZERO    key 

Changes value of blinking digit. 

 PRINT  key 

Stores and proceeds to step 12. 

 CAL  key 

Cancels and proceeds to step 12. 

Step 12  Press the    CAL    key to return to weighing mode.GXM-06 Analog Voltage Output 

The analog voltage output (GXM

-

04) can convert the specified weighing value to analog 

voltage between 0 and 1 V and output it. 

Summary of Contents for GX-M Series

Page 1: ...GXM 04 GXM 06 COMPARATOR OUTPUT ANALOG VOLTAGE OUTPUT 1WMPD4003912A GX M Series Options ...

Page 2: ...oduced transmitted transcribed or translated into any language in any form by any means without the written permission ofA D Company Ltd The contents of this manual and the specifications of the instrument covered by this manual are subject to change for improvement without notice ...

Page 3: ...ications of comparator Output 11 6 2 Using the Comparator Output 13 6 3 ComparisonMethodSelection ToleranceRangeandCenter ValueorThresholdValues 14 6 4 Setting Threshold Values 15 6 5 Setting Center Value and Tolerance Range 17 7 GXM 06Analog Voltage Output 19 7 1 Specifications ofAnalog Voltage Output 19 7 2 Function Table ofAnalog Voltage Output 20 7 3 Voltage Output Selection 22 7 4 Voltage Out...

Page 4: ...puts HH HI OK LO LL of the comparison output and READY output to indicate the status of the balance 3 stage comparison or 5 stage comparison can be selected The GXM 04 can sound a buzzer according to the comparison result The GXA 04 is equipped with external control input terminals that can make the RE ZERO and PRINT key operations An optional foot switch AX SW137 PRINT AX SW137 RE ZERO can be use...

Page 5: ...to the following procedure The installation procedure of the GXM 04 and GXM 06 is the same Step1 Remove the two screws from the panel and remove the RS 232C USB board from the balance Step2 Pull gently and remove the two cables from theRS 232C USB board Step3 Insert the two cables into the sockets of the GXM 06 board Step4 Insert the rubber packing and plug the GXM 06 into the balance Screw the pa...

Page 6: ... terminal Step3 Cover the terminal with the top metal case and bottom metal case Step4 Fasten wires with the top metal case Step5 Insert the assembled parts into the rubber sleeve Rubber sleeve Top metal case Terminal Bottom metal case Solder wires and terminal Fastenwirestightly Cable Part to secure wires ...

Page 7: ...ach item stores a parameter The last displayed parameter is in effect After the PRINT key is pressed these updated parameters are applied to operations of the balance Example of settings and menu structure Example Set to Center value using sample weight input of Input method in Comparator Caution The balance may not function properly depending on the settings and the operating environment Be sure ...

Page 8: ...m zero 3 Comparison excluding 20 digits counts from zero 4 Comparison excluding 50 digits counts from zero 5 Comparison excluding 100 digits counts from zero CP P Polarity 0 Comparison of positive values only 1 Comparison of negative values only 2 Comparison of both polarities CP r Comparison results 1 Not added Select whether or not to add the comparison results to the output data 2 Added CP in I...

Page 9: ...5 stage comparison is selected CP Hi Upper limt CP lo Lower limit CP LL Second lower limit CP ref Center value Refer to explanation of 6 GXM 04 Comparator output Referencevaluewhenusingthedatainputmethod 2 or 3 isselected is displayed when 5 stage comparison is selected Tolerance range from center value Second tolerance range from center value indicates factory settings Digit is the unit of minimu...

Page 10: ...Binterface Aout Analog output An Analog output mode 0 2 digit output mode Displayed only when the GXM 06 is connected 1 3 digit output mode 2 Net full scale output mode 3 Gross full scale output mode 5el Output digit selection 0 First digit Displayed only when the GXM 06 is connected 1 Second digit 2 Third digit 3 4th digit 4 5th digit 5 6th digit 6 7th digit AP fnc Application Refer to instructio...

Page 11: ...act resistance 20 Ω Outputconnector Compatibleplug 8pinminiDINconnector Female Accessory E8 200J 100 or equivalent Externalcontactinputconnector Compatibleplug φ3 5mm3poleministereo jack MP 013LC or equivalent parts Panel External View 8 7 6 3 Comparator output connector 4 External contact input x2 2 3 1 5 2 1 RS 232C connector Example of contact input terminal Comparator output circuit Pin assign...

Page 12: ...an connect to either input Straight cable About 5V Using External Contact Input When 1 pin and 3 pin or 1 pin and 2 pin of lead wire of optional foot switch AX SW137 PRINT AX SW137 RE ZERO is shorted 100 ms or longer it can be operated the same as the PRINT key and RE ZERO key on the panel ...

Page 13: ...merical input Another is the input using sample weight Balance can sound a buzzer according to the comparison result That can be specified at Comparator buzzer in the function table When weighing value is not displayed in performing RE ZERO etc READYcontact becomes open When weighing value is displayed READYcontact becomes short 3 stage comparison result 3 stage comparison display contact output W...

Page 14: ...tep3 Press the PRINT key to enter Comparator Step4 Press the RE ZERO key several times to display Comparison always Step5 Press the SAMPLE key to display Comparison method Step6 Press the RE ZERO key to display 5 stage comparison Step7 Press the SAMPLE key to display Near zero Step8 Press the RE ZERO key several times to display Comparison excluding 5 digits counts from zero Step9 Press the SAMPLE...

Page 15: ...old value is displayed the current parameters are displayed all digits blinks Numerical input of threshold values If Step6 If changing the parameters press the RE ZERO key to store the following keys SAMPLE key Moves blinking digit RE ZERO key Changes value of blinking digit MODE key Changes polarity PRINT key Stores and proceeds to step 7 CAL key Cancels and proceeds to step 7 If the current para...

Page 16: ...ond upper limit upper limit lower limit and second lower limit can be specified using RS 232C commands from peripherals Example command HH 1000 000 g second upper limit HI 0800 000 g upper limit LO 0600 000 g lower limit LL 0400 000 g second lower limit space ofASCII code 20 h ...

Page 17: ... parameter is displayed all digits blinking If parameter does not need to be changed press the PRINT key or CAL key and proceed to step 6 If the parameter is to be changed press the RE ZERO key and change it using the following keys SAMPLE key Moves blinking digit RE ZERO key Changes value of blinking digit MODE key Changes polarity PRINT key Stores and proceeds to step 6 CAL key Cancels and proce...

Page 18: ...s and proceeds to step 11 Step11 Press the PRINT key while displaying The current parameter is displayed If the second tolerance range is to be changed change it using the following keys Input the second tolerance range with center value assumed to be 100 SAMPLE key Moves blinking digit RE ZERO key Changes value of blinking digit PRINT key Stores and proceeds to step 12 CAL key Cancels and proceed...

Page 19: ... or equivalent Pin connections Output 4 pin GND 3 pin Output range 0 V to 1 V With the slide switch set to 0V 0 2 V to 1 V With the slide switch set to 0 2V Input impedance of 10 kΩ or greater the device connected Panel External View 3Analog GND Analog output connector 4Analog voltage output Volume for fine adjustment 2 1 Output range RS 232C connector Output circuit Pin assignments and wire color...

Page 20: ...ng the RE ZERO key 3 Gross full scale output mode Outputs 0 000 V when the gross weight is zero Outputs 1 000 V when the gross weight is full scale Tare operation using the RE ZERO key will not affect the output Note If tare is extremely light tare operation might change the zero point Thus affecting the output 5el Output digit selection Specify minimum digit to be output at mode When is set to or...

Page 21: ...If an 2 or an 3 Model Full scale GX 8202M GX 8202MD GF 8202M GF 8202MD 8 kg GX 10202M GF 10202M 10 kg GX 12001M GF 12001M 12 kg GX 22001M GF 22001M 22 kg GX 32001M GX 32001MD GF 32001M GF 32001MD 32 kg Example When the GX 10202M displays 1 kg the voltage output of is 0 1 V Using 0V of the slide switch 1 000V X 0 100V Caution Full scale of the full scale output mode indicates the full scale values ...

Page 22: ...d a voltmeter voltage output can be finely adjusted Procedure Step1 Turn off the display using the ON OFF key Voltage output becomes zero point status Adjust the ZERO volume so that the voltmeter indicates 0 000 V when the slide switch is set to 0V Adjust the ZERO volume so that the voltmeter indicates 0 200 V when the slide switch is set to 0 2V Step2 While pressing and holding the PRINT and SAMP...

Page 23: ...2V During the re zeroing operation in 3 digit output mode 0 V is output or 0 2 V if the slide switch is set to 0 2V Step3 When E weighing value is negative overload is displayed 0 V is output or 0 2 V if the slide switch is set to 0 2V Step4 When E weighing value is positive overload is displayed the following voltages are output Example Voltage output when E is displayed if the slide switch is se...

Page 24: ...treet Thebarton South Australia 5031 AUSTRALIA Telephone 61 8 8301 8100 Fax 61 8 8352 7409 A D KOREA Limited 한국에이 엔 디 주 서울특별시 영등포구 국제금융로6길33 여의도동 맨하탄빌딩 817 우편 번호 07331 817 Manhattan Bldg 33 Gukjegeumyung ro 6 gil Yeongdeungpo gu Seoul 07331 Korea 전화 82 2 780 4101 팩스 82 2 782 4264 OOO A D RUS OOO ЭЙ энд ДИ РУС 121357 Российская Федерация г Москва ул Верейская дом 17 Business Center Vereyskaya Plaza...

Reviews: