Assembling the Kit
(On the balance without a glass breeze break installed)
(1) Remove the weighing pan from the balance.
(2) Place the density pan stand on the balance.
(3) Set the beaker stand so that it does not touch
the density pan stand. Set the beaker stand
spacers under the beaker stand legs so that the
legs fit into the grooves of the spacers.
(4) Attach the thermometer clamp to the beaker
and insert the thermometer in the
thermometer clamp holes.
(5) Pour a liquid (distilled water, etc.) - whose
specific gravity is known - into the beaker
and place the beaker on the beaker stand.
(6) Place the density pan on the density pan
stand.
(7) Adjust the amount of the liquid so that the
sample is about 10 mm below the surface
of the liquid with the sample placed on the
lower pan (in the liquid).
(8) When the value displayed on the balance
becomes stable, press the RE-ZERO key
to reset the displayed value to zero.
Preparations
for
measurement are now
complete.
Note
If the glass breeze break is not available and the display is hard to stabilize, set
“Condition (
Cond
) ” of “Environment/Display” (
ba5fnc
) in the function table to a
greater value.