background image

GP SERIES 

GP-12K/GP-20K /GP-30K/GP-32K 

GP-40K/GP-60K/GP-61K 

GP-100K/GP-102K 

GP-30KS/GP-32KS   

GP-60KS/ GP-61KS/GP-100KS 

 

Precision Balance 

INSTRUCTION MANUAL 

WM+PD4000273D 

Summary of Contents for GP-61KS

Page 1: ...GP SERIES GP 12K GP 20K GP 30K GP 32K GP 40K GP 60K GP 61K GP 100K GP 102K GP 30KS GP 32KS GP 60KS GP 61KS GP 100KS Precision Balance INSTRUCTION MANUAL WM PD4000273D...

Page 2: ...duced transmitted transcribed or translated into any language in any form by any means without the written permission of A D Company Ltd The contents of this manual and the specifications of the instr...

Page 3: ...tic Response Adjustment Self Check Function 23 6 2 Manual Response Adjustment 24 7 CALIBRATION 25 7 1 Calibration Group 25 7 2 Automatic Self Calibration Calibration due to changes in temperature 26 7...

Page 4: ...and Storage Mode 72 12 PROGRAMMABLE UNIT 73 13 DENSITY MEASUREMENT 74 RS 232C serial interface 14 I O UNIT SPECIFICATIONS Standard 77 14 1 RS 232C External Contact Input 77 15 CONNECTION TO PERIPHERAL...

Page 5: ...IP65 allows washing with water Built in Calibration Weight hereinafter referred to as internal mass allows easy calibration adjustment and maintenance of the balance Automatic Self Calibration using t...

Page 6: ...nce has a swing arm to adjust the display position for optimum viewing Separate display type balances GP 30KS 32KS 60KS 61KS 100KS are also available Windows Communication Tools WinCT allows easy comm...

Page 7: ...ovide reasonable protection against interference when equipment is operated in a commercial environment If this unit is operated in a residential area it may cause some interference and under these ci...

Page 8: ...6...

Page 9: ...terial to be used for transporting the balance in the future The packing contents depend on the balance model See the illustrations to confirm that everything is contained Standard type GP 12K 20K 30K...

Page 10: ...Install the balance as follows 1 Refer to 3 PRECAUTIONS for installing the balance 2 Adjust the leveling feet to level the balance Confirm it using the bubble spirit level 3 Confirm that the adapter t...

Page 11: ...and as level as possible Level the balance by adjusting the leveling feet and confirm it using the bubble spirit level Ensure a stable power source when using the AC adapter Plug in the AC adapter and...

Page 12: ...nterior free of dust and foreign materials 3 3 After Use Avoid mechanical shock to the balance Do not disassemble the balance Contact the local A D dealer if the balance needs service or repair Do not...

Page 13: ...on will interrupt the operation and turn the display OFF In the weighing mode turns the minimum weighing value ON and OFF In the counting or percent mode enters the sample storing mode Enters the func...

Page 14: ...ey to activate the precision range again Below is a description of how the Smart Range actually functions using the GP 102K Precision range 61 kg 1 g Standard range 101 kg 10 g 1 Press the RE ZERO key...

Page 15: ...s see the table below Name unit mode Abbrev Display Function table Storing mode Conversion factor 1 g Gram g G 1 g kilogram kg 1000 g Counting mode PCS Percent mode Ounce Avoir OZ 28 349523125 g Pound...

Page 16: ...5 39223 9 0 1 26363 5 0 5 Tael HK general Singapore 317 465 555 565 820 120 1613 780 0 005 1084 68 0 02 Tael HK jewelry 320 605 561 060 828 235 1629 750 0 005 1095 40 0 02 Tael Taiwan 320 000 560 000...

Page 17: ...975 0 005 Tael Taiwan 826 65 0 05 162 665 0 005 Tael China 992 00 0 05 195 200 0 005 Tola India 2657 8 0 1 522 99 0 01 Messghal 6613 5 0 5 1301 35 0 05 GP 102K Standard range Precision range Unit Capa...

Page 18: ...he SAMPLE key several times to display Unit 3 Press the PRINT key to enter the unit selection mode 4 Specify a unit or mode in the order to be displayed using the following keys SAMPLE key To sequenti...

Page 19: ...ends on the balance model 2 Place a sample on the pan or in the container 3 Wait for the stabilization indicator to be displayed Read the value 4 Remove the sample and container from the pan Notes To...

Page 20: ...l the weight tare The number specified in step 3 appears e g 25 0 pcs is displayed if 25 is selected in step 3 5 Place the number of samples specified on the pan In this example 25 pieces 6 Wait for t...

Page 21: ...ded Try to add the same number of samples as displayed 9 The balance re calculates the unit mass while the processing indicator is blinking Do not touch the balance or samples on the pan until the pro...

Page 22: ...iner 5 Press the PRINT key to store the reference mass The balance displays 100 00 The decimal point position depends on the reference value The reference mass stored even if the AC adapter is removed...

Page 23: ...set to 0 Selecting the unit 1 Press the MODE key to select a unit to be used for accumulation Note While the accumulation function is in use unit selection using the MODE key is not available Setting...

Page 24: ...zero 2 Place a sample on the pan The weight value is displayed 3 Press the PRINT key The weight value is added to the total and output The accumulation number at the upper left of the display increase...

Page 25: ...te is FAST or 10 times second when the response rate is MID or SLOW change the Display refresh rate 5pd parameter of Environment Display ba5fnc in the function table For details refer to 9 FUNCTION TA...

Page 26: ...2 Manual Response Adjustment This function manually updates the response adjustment Operation 1 Press and hold the MODE key until RESPONSE is displayed then release the key And then press the MODE ke...

Page 27: ...drafts to affect the balance during calibration To output the data for GLP using the RS 232C interface set GLP output info of Data output dout For details refer to 9 FUNCTION TABLE Time and date are a...

Page 28: ...n To maintain the calibrated state keep the weighing pan clear while not in use The displays shown below are related to the automatic self calibration Indicates that the balance detects a change in am...

Page 29: ...arameter of the function table is set to 1 or 2 the balance displays glp and outputs Calibration Report using the RS 232C interface or stores the data in memory For details on the calibration report f...

Page 30: ...bration weight value stored in the balance proceed to step 5 4 Specify the calibration weight value as follows SAMPLE key To switch the display condition to All of the segments blinking calibration we...

Page 31: ...Remove the weight from the pan 8 If the GLP output info parameter of the function table is set to 1 or 2 the balance displays glp and outputs Calibration Report using the RS 232C interface or stores t...

Page 32: ...ghts is shown on page 25 press the SAMPLE key and proceed to step 4 If you use the target weight value stored in the balance proceed to step 5 4 Specify the target weight value as follows SAMPLE key T...

Page 33: ...n and press the PRINT key The balance measures the target weight and displays the measured value Do not allow vibration or drafts to affect the balance 7 The balance displays end Remove the weight fro...

Page 34: ...eight The corrected mass value is maintained in non volatile memory even if the AC adapter is removed The internal mass value is corrected as follows Model Correction reference value Correction range...

Page 35: ...the SAMPLE key to enter the function table and release the key when ba5fnc is displayed 7 Press the SAMPLE key several times until C5in is displayed then release the key 8 Press the PRINT key to enter...

Page 36: ...rations Switches Operation 1 Press the ON OFF key to turn off the display 2 While pressing and holding the PRINT key and the SAMPLE key press the ON OFF key The balance displays p5 3 Press the PRINT k...

Page 37: ...Function switch settings Liquid density and temperature in the density mode Note Be sure to calibrate the balance after initialization Operation 1 Press the ON OFF key to turn off the display 2 While...

Page 38: ...ing data for Data memory and 1 minute for Interval time 9 2 Display and Keys Display Key Description The symbol indicates that the parameter displayed is in effect When pressed and held in the weighin...

Page 39: ...37 9 3 Details of the Function Table P...

Page 40: ...38...

Page 41: ...39 Caution The balance may not transmit the data completely at the specified refresh rate depending on the baud rate or data added to the weighing data such as time date and ID number...

Page 42: ...weighing 5t b 2 This parameter ignores slight fluctuation of a weight value Used to prevent a weight value from drifting due to vibration or drafts Note With Hold function Hold set to ON 1 this item...

Page 43: ...o display the weighing mode Used when the balance is built into an automated system 30 minute warm up is necessary for accurate weighing Auto display OFF poff When the AC adapter is connected and no o...

Page 44: ...o print mode A reference zero dout ap p Auto print polarity dout ap b Auto print difference Example For weighing each time a sample is placed and removed with ar d set to 1 to adjust zero after the da...

Page 45: ...his format consists of fifteen characters excluding the terminator A header of two characters indicates the balance condition The polarity sign is placed before the data with the leading zeros If the...

Page 46: ...only for negative data The weighing data uses spaces in place of the leading zeros The character length of this format changes dependent upon the unit NU numerical format 5if type 4 This format output...

Page 47: ...er can be changed in Clock Cl adj Outputs the year in four digit format Time dout 5 td 1 or 3 Outputs time in 24 hour format Tare value When the tare value in memory is recalled the tare value is outp...

Page 48: ...46 9 8 Data Format Examples...

Page 49: ...47 Note When Pound Ounce is selected the data is output with the unit of ounce oz...

Page 50: ...ing the time 4 The current time is displayed with all the digits blinking When the time is correct and the date does not need to be confirmed press the CAL key and proceed to step 8 When the time is c...

Page 51: ...the following keys SAMPLE key To select the digits to change the value The selected digits blink RE ZERO key To increase the value by one MODE key To decrease the value by one PRINT key To store the n...

Page 52: ...e SAMPLE key until ba5fnc of the function table is displayed then release the key 2 Press the SAMPLE key several times to display Cp fnc 3 Press the PRINT key 4 Press the RE ZERO key several times to...

Page 53: ...ress the SAMPLE key several times to display Cp fnc 3 Press the PRINT key 4 Press the RE ZERO key several times to display Cp 4 5 Press the SAMPLE key to display Cp in 6 Press the RE ZERO key several...

Page 54: ...used to set the tolerance value 10 Press the CAL key to exit the comparator function and return to the weighing mode Notes When Pound Ounce is selected as a weighing unit enter the values in ounces fo...

Page 55: ...s the RE ZERO key to set the display to zero 2 Place a sample on the pan The balance performs a comparison using the specified comparison values and displays the comparison results HI OK or LO 3 Each...

Page 56: ...ion using an external weight Calibration test report of the calibration test using an external weight Title block and End block for the weighing data Calibration and calibration test data can be store...

Page 57: ...e AD 8121B Printer To output the report to a personal computer using the RS 232C interface set the GLP output info parameter to 2 If the time and date are not correct set the correct time and date in...

Page 58: ...ng an external weight When the setting is info 1 When the setting is info 2 Calibration test report using an external weight Calibration test does not perform calibration When the setting is info 1 Wh...

Page 59: ...ion If the data memory function is used the Title block and End block can not be output Operation 1 With the weighing data displayed press and hold the PRINT key then release the key S5tart is display...

Page 60: ...a memory refer to 9 FUNCTION TABLE For weighing data the data contents to be stored and the storage capacity depend on the Time Date output 5 td parameter setting Releasing Clr If a different type of...

Page 61: ...to 2 2 Set the Time Date output 5 td parameter as necessary 3 The storing operation depends on the Data output mode prt parameter setting Four types of operating modes are available to store data Key...

Page 62: ...Not used Interval memory mode prt 3 Not used data 2 int 0 8 Set each item depending on the situation as follows No d no 0 No 5 td 0 Data number Yes d no 1 Time only 5 td 1 No 5 id 0 Date only 5 td 2 I...

Page 63: ...firm that the Serial interface 5if parameters are set properly For details refer to 9 FUNCTION TABLE and 15 CONNECTION TO PERIPHERAL EQUIPMENT 1 Press and hold the PRINT key until reCall is displayed...

Page 64: ...y Transmitting the memory data Confirm that the Serial interface 5if parameters are set properly For details refer to 9 FUNCTION TABLE and 15 CONNECTION TO PERIPHERAL EQUIPMENT and also confirm that t...

Page 65: ...lect pcs counting mode If Clr appears blinking in the upper left of the display delete the data in memory Note If the counting mode can not be selected refer to 4 WEIGHING UNITS 3 Press and hold the P...

Page 66: ...shown to the right The digit to be changed blinks Note ACAI can not be used on the unit mass stored using the digital input mode Use the following keys to store a unit mass in digital input mode SAMPL...

Page 67: ...y is not available 2 Set the Data memory data parameter to 4 If Clr appears blinking in the upper left of the display delete the data in memory as described in 11 1 Notes on Using Data Memory 3 Press...

Page 68: ...sample is placed on the pan to store the upper and lower limit values Notes Pressing the CAL key will interrupt the operation and the balance will return to the upper and lower limit values confirmat...

Page 69: ...ey to confirm the selection The balance returns to the weighing mode with the selected upper and lower limit values ready for use To cancel the selection press the CAL key The balance returns to the w...

Page 70: ...il the balance enters the tare value confirmation mode The tare value last selected is displayed as shown below All the digits blinking 4 Select the tare number to be used using the following keys RE...

Page 71: ...RO key to set the display to zero 3 Place a tare a container on the pan 4 Press the PRINT key to store the tare value Remove the tare from the pan Digital input mode In the digital input mode the tare...

Page 72: ...to select the value Each time the MODE key is pressed the displayed value changes as follows t03 t04 t05 t20 t01 4 Press the PRINT key to confirm the selection The balance returns to the weighing mode...

Page 73: ...ast selected appears with all the digits blinking 2 Press the MODE key to select the value Comparator settings when the Data memory data parameter is set to 4 Each time the MODE key is pressed the dis...

Page 74: ...data number by one Unit mass when the Data memory data parameter is set to 1 Each time the MODE key or RE ZERO key is pressed the displayed value changes as follows p03 p04 p05 p20 p01 Comparator set...

Page 75: ...et the coefficient Confirming the coefficient 4 The current coefficient is displayed with the first digit blinking When it is not to be changed press the CAL key and proceed to step 6 When it is to be...

Page 76: ...rmula Where A Density of a sample A Weight value of a sample in air B Weight value of a sample in liquid B Density of a liquid Prior to measurement Changing the function table Prior to measurement cha...

Page 77: ...parameter to set 3 Remove the sample from the upper pan Notes If an unstable value a negative value or e error is displayed the PRINT key is disabled The SAMPLE key can not be used to change the minim...

Page 78: ...e the condition of step 1 2 Selecting a parameter to set The relation between the water temperature and density is shown below Enter the density directly ldin 1 The density currently set unit g cm3 fa...

Page 79: ...put Input Same as the PRINT key 19 RE ZERO External contact input Input Same as the RE ZERO key Others No connection RS 232C The balance is a DCE device Connect the balance to a personal computer DTE...

Page 80: ...pin 7 or pin 19 to pin 7 for more than 100 ms the same operation as performed by pressing the RE ZERO key or the PRINT key will be performed Use of example The external input connector AX HDB 25P CTF...

Page 81: ...r is used dout 5 td 0 3 Outputs the time and date as necessary dout 5 id 0 1 Outputs the ID number as necessary dout at f 0 1 Uses auto feed as necessary 5if type 1 DP format When data is transmitted...

Page 82: ...demarks of Microsoft Corporation 15 4 Using the WinCT software the balance can do the following 1 Analyzing the weighing data and the statistics with RsKey The weighing data can be input directly into...

Page 83: ...mass value CN mm Recalls the upper lower limit value in memory mm indicates 01 20 CN Requests the upper lower limit value code number of the selected value H I g Sets the upper limit values Command e...

Page 84: ...eleased using the CAL command CAL command Calibration command ON command Display ON command P command Display ON OFF command R command RE ZERO command When a communication error has occurred due to ex...

Page 85: ...hecking contact the local A D dealer for repair Checking that the balance performs properly Check the balance performance using the self check function as described in 6 1 Automatic Response Adjustmen...

Page 86: ...CHECK FUNCTION about adapting the balance to the environment To return to the weighing mode press the CAL key Out of the setting range The data to be stored is out of the setting range EC E16 Interna...

Page 87: ...nse adjustment unstable error The automatic response adjustment can not be performed because the weight value is unstable Check the ambient conditions such as breeze vibration and magnetic fields also...

Page 88: ...racters in a command Confirm the command EC E06 Format error A command includes incorrect data e g The data is numerically incorrect Confirm the command EC E07 Parameter setting error The received dat...

Page 89: ...C with Windows Communication Tools WinCT External calibration weight 5 kg 10 kg 10 kg 20 kg 20 kg 30 kg 20 kg 30 kg 40 kg Weighing pan 384 x 344 mm Dust and water protection Complying with IP65 Sensor...

Page 90: ...ernal calibration weight 30 kg 40 kg 50 kg 60 kg 60 kg 80 kg 100 kg Weighing pan 384 x 344 mm 386 x 346mm Dust and water protection Complying with IP65 Sensor used Super Hybrid Sensor SHS AC adapter C...

Page 91: ...89 20 EXTERNAL DIMENSIONS Standard Type GP 12K 20K 30K 32K 40K 60K 61K 100K 102K Note The size in parentheses is for the GP 100K 102K Unit mm...

Page 92: ...90 Separate Display Type GP 30KS 32KS 60KS 61KS 100KS Note The size in parentheses is for GP 100KS Unit mm...

Page 93: ...tandard RS 232C serial interface So RS 232C is not available for use OP 07 Extension cable 5 m Available as factory option only when ordered with a GP balance Applicable balance models GP 30KS 32KS 60...

Page 94: ...be included in the weighing data Normally refers to an operation of placing a container and setting the display to zero Mode Balance operational function Re zero To set the display to zero GLP Good L...

Page 95: ...ba5fnc 37 bep 37 bep 37 bep_ 37 bp5 39 btpr 39 Bubble sprit level 7 8 C C5 in 39 Cal e 84 Cal e 84 Cl adj 37 Cond 37 40 Cp 37 Cp fnc 37 Cp Hi 37 Cp in 37 Cp lmt 37 Cp ref 37 Cp b 37 Cp lo 37 Cp r 37 C...

Page 96: ...cent mode 20 Permit 34 Power consumption 87 88 PRINT key 11 16 Printer bracket 91 Prior notice indicator of automatic self calibration 11 Processing indicator 11 R rng 37 41 Rechargeable battery 91 Re...

Page 97: ...95 MEMO...

Page 98: ...96 MEMO...

Reviews: