52
Title block and end block
When a weight value is recorded as the GLP data, “Title block” and “End block” are inserted at the
beginning and at the end of a group of weight values, in the GLP report.
Note
To output the report to an AD-8121B , use MODE 3 of the AD-8121B.
Operation
1 With the weight data displayed, press and hold the PRINT key until
S5tart
is displayed.
The “Title block” is output.
2 The weight data is output according to the parameter setting of the data output mode.
3 Press and hold the PRINT key until
recend
is displayed. The “End block” is output.
When the setting is “
info 1
”:
When the setting is “
info 2
”: