32
KF format
5if type 2
This is the Karl-Fischer moisture meter format and is used when the peripheral equipment can only
communicate using this format.
This format consists of fourteen characters excluding the terminator.
This format has no header characters.
The polarity sign is placed before the data, with spaces in place of leading zeros, if the data is
not zero or overloaded.
This format outputs the unit only for a stable value.
MT format
5if type 3
A header of two characters indicates the balance condition.
The polarity sign is used only for negative data.
The weight data uses spaces in place of the leading zeros.
The character length of this format changes dependent upon the unit
NU (numerical) format
5if type 4
This format outputs only numerical data.
This format consists of nine characters excluding the terminator.
The polarity sign is placed before the data with the leading zeros. If the data is zero, the plus
sign is used.
CSV format
5if type 5
Separates the data of A&D standard format and the unit by a comma (,).
Outputs the unit even when the data is overloaded.
+ 1 . 2 7 8 c t
CR LF
Data
Unit
Terminator
c t
Stable value
Unstable value
S 1 . 2 7 8 c t
CR LF
Header
Unit Terminator
Data
+ 0 0 0 1 . 2 7 8
CR LF
Data
Terminator
CR LF
O L , + 9 9 9 9 9 9 9 E + 1 9 , c
t
S T , + 0 0 0 1 . 2 7 8 , c t
CR LF
ST,+0001.278, c t<CR><LF>
Weight data
Summary of Contents for FX-1200GD