64
14. KEY OPERATION
Test mode
To enter the test mode, press and hold [7] and [8], and turn on the power.
“CHECK_MODE” will be displayed and the program version number will be
printed.
To return to the PRINT mode, turn off the power and turn on the power again.
Initialization
To initialize, press and hold [7] and [9], and turn on the power.
・
Restores the function settings to the default settings.
・
Initializes the printing format and interval timer setting.
・
Clears 6-digit code number data, subtotal/grand total, data buffer contents.
Release the keys when “INIT_ALL” is displayed. After initialization, the printer
returns to the PRINT mode.
Clearing data in
random printing
(
F3=1/F9=1
)
Press and hold [CAN.], and press [PRINT] to print “* CANCEL” and clear the
data added last in memory.
・
Only one data that is added last can be cleared.
Clearing data
buffer in batch
printing
(
F3=1/F9=2
)
Press and hold [CAN.], and press [PRINT] to print “* CANCEL” and clear the
data in the data buffer.
Note that the total value will not be cleared as the data received is added.
Clearing
grand total/Lot
printing data
Press and hold [CAN.], and press [G.TTL] to print “* CANCEL” and clear
grand total and lot printing contents.
・
Clears both the grand total data and the lot printing data.
・
When clearing the grand total data, the batch printing data will not be
cleared.