7. CALIBRATION 

The AD-4530 measures voltage signals from sensors and displays the 

values. Calibration is performed so that it performs correctly. 

 

The decimal point (CF01), minimum scale value (CF02) and rated 

capacity (CF03) are set using the function mode. 

The zero point input voltage (CF04), the span input voltage (CF05) 

and the display value for the span input voltage (CF06) are set using 

calibration mode. 

Calibration setting by function mode is also available. (Digital 

calibration) 

 

* During calibration, maintain a stable environment or a calibration 

error may occur. 

* You can confirm stabilization by confirming that the HOLD LED is on. 

* The decimal point blinks to indicate that a number is being displayed, 

not a weighing value.

 

 

7.1. Calibration Modes 

With the weighing mode, press      and       to enter calibration 

mode. 

 

 

Enter zero point calibration mode. 

   

 

Return to the weighing mode. 

 

 

7.1.1.

 

Zero point calibration mode

 

With nothing on the load cell, wait for 

it to stabilize, and then press            . 

     

Perform zero point calibration, and 

proceed to span calibration mode. 

 

 

Cancel zero point calibration, and 

proceed to span calibration mode. 

   

Hold to display the mV/V of the zero 

point. 

 

 

 

 

 

 

 

7.1.2.

 

Span calibration mode

 

Add the actual load for span calibration to the load 

cell, and then input the value to be displayed when 

the load is loaded.

 

After stabilization, press      . 

 

Select the digit to be changed. 

 
Increase the value of the digit to be changed. 

 
Change the polarity. 

 

Perform span calibration, and 

proceed to storing mode. 

 

Cancel span calibration, and 

proceed to storing mode. 

 

* After the span calibration, the AD-4530 displays the mV/V value of the span 

calibration for 3 seconds, and then proceeds to the storing mode. 

 

7.1.3. Storing calibration mode

 

Store the calibration zero, span and displayed 

value acquired. 

When calibration was not performed, data is not 

stored.  

 

Store the data acquired, and return to 

weighing mode. 

 

Do not store the data acquired, and 

return to weighing mode. 

  

7.2. Calibration Error 

Display Cause 

Remedy 

C?e2 

Voltage at zero point 

calibration was too far in the 

positive direction. 

C?e3 

Voltage at zero point 

calibration was too far in the 

negative direction. 

Confirm the rating 

and connection of the 

load cell. 

 

Display Cause 

Remedy 

C?e4 

The mass value exceeds the 

weighing capacity. 

Use a proper mass. 

C?e5 

The mass value is less than 

the minimum scale value. 

Use a proper mass. 

C?e6 

The load cell sensitivity is 

insufficient. 

Confirm the load cell 

connection. Use a proper 

mass. 

C?e7 

Voltage at span calibration is 

less than the zero point. 

Confirm the load cell 

connection. 

C?e8 

When adding a load of the 

weighing capacity, the load cell 

output voltage is too high. 

Use a load cell with a 

greater weighing capacity 

or set a smaller weighing 

capacity value. 

 

8. FUNCTION MODE 

Use the function mode to set various functions and data. The set values are 

saved in non-volatile memory and are maintained even if the power is 

switched off. 

 

8.1. Structure of Functions 

The first 2 digits of the Function No. are the function group. The last 2 digits of the 

Function No. are the function item. 

 

CF Calibration 

function 

F0 Basic 

function 

F1 Comparator 

function 

 

Use this function to set the comparator operation. 

F2 

Analog output function 

 

Use this function to set the output value of the analog voltage 

output and analog current output. 

F3, F4  Serial communication function 

 

Use this function to set the RS-232C and RS-485. 

*   Set the zero point input voltage (CF04), the span input voltage (CF05) and 

the display value for the span input voltage (CF06) in calibration mode. 

* Set the upper limit value (F101) and lower limit value (F102) in 

comparator mode. 

 

When setting a function, the decimal point blinks to indicate that a 

number is being displayed, not a weighing value. 

 

8.2.Key operation 

In weighing mode, press       and       to enter the function selection 

mode. 

 

 

 

8.2.1. Function selection mode 

 

Select the function group. (First 2 digits) 
 
Select the function item. (Last 2 digits) 
 
Enter the setting changing mode. 
 

Save the setting in non-volatile memory, and then return to 
weighing mode. 

 

8.2.2. Setting changing mode (Two methods) 

 

 

Parameter selection method (All digits blinking) 

 

Change the parameter. 
 
Enter the setting and return to function selection mode. 
 
Cancel the setting and return to function selection mode. 

 

 

 

Digital input method (Change the blinking digit only) 

 

Move the digit to be changed to the right. 
 
Change the value of the blinking digit. 
 
Change the polarity. 
 

Enter the setting and return to function selection mode. 

 

Cancel the setting and return to function selection mode. 

 

 

8.3. Function Items 
8.3.1. Calibration (C function) 

Function No.
SETTING range

Function

Description 

Default value 

Setting type 

Cf01 

0 to 3 

Decimal point 

position 

Decimal point position of the weighing 

value 

0

:

0000 

1

:

000.0 

2

:

00.00 

3

:

0.000

 

P

 

Cf02 

1 to 50 

Minimum 

scale value

Minimum division (d) of the weighing value 

1

:1           

10

:10 

2

:2           

20

:20 

5

:5           

50

:50 

P

 

Cf03 

1 to 9999 

Maximum 

weighing 

capacity 

Weighing is possible up to the value of 

this setting plus 8 d (8 minimum 

divisions) 

Decimal point position depends on

 

Cf01

 . 

7000 

Cf04 

-7.000 

 to 7.000 

Input voltage 

of zero point

Input voltage from the load cell at zero 

point 

(Unit: mV/V) 

0.000 

Cf05 

0.001 

 to 9.999 

Input voltage 

of span 

Input voltage from the load cell at span 

(measurement point - zero point) 

(Unit: mV/V) 

7.000 

Cf06 

-9999 

 to 9999 

Display value 

for input 

voltage of span

Display value for span (measurement 

point - zero point) 

Decimal point position depends on

 

Cf01

 . 

7000 

Cf07 

0 to 100 

Zero 

adjustment 

range 

Range to enable zero adjustment by the 

ZERO key 
Expressed in percent of the weighing 

capacity with the calibration zero point 

as the center 

100 

Cf08 

0.0 to 5.0 

Zero tracking 

time 

Performed in combination with zero 

tracking width. (Unit: second) 
When 

0.0

, zero tracking is not performed.  

0.0 

Cf09 

0.0 to 9.9 

Zero tracking 

width 

Performed in combination with zero 

tracking time. (Unit: 0.1 d) 
When 

0.0

, zero tracking is not performed.  

0.0 

Cf10 

0 to 2 

Power on zero

Digital zero when turning power on 

0

: Digital zero off 

1

: Perform digital zero again 

2

: Use state when the power was turned 

off 

Cf11 

1 to 2 

Zero operation

1

: On with              , Off with

   

     

+

 

   

2

: On or off with 

     

+

 

     

 

 

8.3.2.

 

Basic Functions

 

Function No.
Setting range

Function

Description 

Default value 

setting type 

f001 

0000 

 to 1111 

Disable key

Each digit of the setting corresponds to a key 

switch. Only available in the weighing mode. 

Key assignment

  0

: Enabled 

0 0 0 0

             

1

: Disabled

 

 

 

 

 

 

 

 

 

 

  

+  

 

0000 

(Binary) 

D

 

f002 

0 to 11 

Digital filter

Cutoff frequency 

0

: Off                     

6

:0.5 Hz 

1

:2.5 Hz             

7

:0.35 Hz 

2

:2.0 Hz             

8

:0.25 Hz 

3

:1.5 Hz             

9

:0.20 Hz 

4

:1.0 Hz      

10

:0.15 Hz 

5

:0.7 Hz      

11

:0.10 Hz 

P

 

f003 

0 to 4 

Hold mode

0

: Off   

 

1

: Sample hold 

2

: Peak hold 

3

: Bottom hold 

4

: Bipolar peak hold 

P

 

f004 

0000 

 to 1111 

Latch 
function 

Function corresponds to an external input 

latch. 

Setting and latch   

assignment

    0

: Off     

0 0 0 0      1

: On

 

Display value latch  

Comparator latch 

Analog output latch   

Serial output latch 

0000 

(Binary) 

* If you want to confirm the weighing value when setting the digital filter (F002), 

press              . 

 

When the weighing value is displayed, the OK LED blinks and the display can 
be set to zero by pressing       . 

 
 Press       to return the setting display. 

 

 

8.3.3. Comparator

 

Function No. 

setting range 

Function

Description 

Default value

setting type

f101 

-9999 to 

9999 

Upper limit 

value 

Upper limit value of comparator. 

Decimal point position depends on

 

Cf01

0000

f102 

-9999 to 

9999 

Lower limit 

value 

Lower limit value of comparator. 

Decimal point position depends on

 

Cf01

0000

f103 

0 to 2 

Comparator

mode 

0

:Off 

1

:On when the measured value is not 

near zero 

2

:Always on 

f104 

-9999 to 

9999

 

Near zero Set the near zero range for Comparator 

mode. 

0000

f105 

1 to 3 

Hysteresis 

mode 

Hysteresis direction 

1

:Upward 2-level judgment 

2

:Upper / lower limit judgment 

3

:Downward 2-level judgment 

f106 

0.0 to 5.0 

Hysteresis 

time 

Set the hysteresis time by units of 0.1 seconds.

When 

0.0

, the hysteresis is not used. 

0.0 

f107 

00 to 99 

Hysteresis 

width 

Set the hysteresis width to a two-digit value. 

When 

00

, the hysteresis is not used. 

99 

 

8.3.4. Analog

 

Function No. 
setting range 

Function

Description 

Default value

setting type

f201 

-9999 to 9999 

0 V 

output 

Measurement value at DAV 0V output. 

Decimal point position depends on

 

Cf01

0000

f202 

-9999 to 9999 

10 V 

output 

Measurement value at DAV 10V output. 

Decimal point position depends on

 

Cf01

1000

f203 

-9999 to 9999 

4 mA 

output 

Measurement value at DAI 4 mA output. 

Decimal point position depends on

 

Cf01

0000

F204 

-9999 to 9999 

20 mA 
output 

Measurement value at DAI 20 mA output. 

Decimal point position depends on

 

Cf01

1000

 

8.3.5. Serial Communication

 

Function No. 
Setting range 

Function Description 

Default value

Setting type

f301 

2.4 to 38.4 

Baud rate 

2.4

: 2400 bps 

4.8

: 4800 bps 

9.6

: 9600 bps 

19.2

: 19200 bps 

38.4

: 38400 bps 

2.4 

 

 

 

 

f302 

7 to 8 

Data bit 

length 

7

:7 bits 

8

:8 bits 

f303 

0 to 2 

Parity 

0

:None 

1

: Odd 

2

: Even 

f304 

1 to 2 

Stop bit 

1

:1 bit

 

2

:2 bits

 

f305 

1 to 2 

Terminator 

1

:CRLF 

2

:CR 

f306 

1 to 2

 

Communicat

ion mode 

1

: Stream 

 

2

: Manual print 

 

f307 

00 to 99 

Model No. 

ID that is added to the serial output

When setting to 

0.0

, the ID is not 

added. 

00 

f401 
f402 
f403 
f404 
f405

 

00 to 7F

 

Unit character1

Unit character2

Unit character3

Unit character4

Unit character5

 

Unit character added to serial 

output. 

Set with the hexadecimal ASCII 

code. 

All characters after 

00

 are ignored. 

00 
00 
00 
00 
00

 

(Hexadecimal)

 

 

8.3.6. Error

 

Display Cause 

Remedy 

ad?e 

The data cannot be 

acquired from A/D 

converter. 

Repair is required. 

eepe 

Correct data cannot 

be read from 

EEPROM. 

Perform initialization. 

If the initialization does not clear 

the error, repair is required.   

Cale 

Calibration data 

error. 

Perform calibration. 

dt?e 

A setting value is out 

of range. 

Check setting values and correct 

if necessary. 

 

Storing mode 

Weighing mode 

Span calibration 