13
AD-4411
7. Calibration
Calibrate the AD-4411 so that it can properly convert the signal from the load cell to a load value.
7.1. Settings required before calibration
Before calibration, the following settings are required.
・
Select the measurement unit (CALF01).
The set unit is only used for external communication. Attach the unit label to the main display.
0: None / 1: g / 2: kg (default) / 3: t
・
Select decimal point position (CALF02).
0:
0 (None) (default)
1:
0.0
2:
0.00
3:
0.000
4:
0.0000
5: 0.00000
・
Set the minimum division (CALF03).
1: 1d (default) / 2: 2d / 3: 5d / 4: 10d / 5: 20d / 6: 50d
・
Set the maximum capacity (CALF04). (Default=999999)
7.2. Digital calibration
After calibration using calibration weights is completed, digital calibration function settings are automatically
updated.
Be sure to make a note of the updated settings so you can re-calibrate without applying actual load if the
AD-4411 is replaced.
・
Load cell input signal at zero calibration (dCAL01)
・
"Load cell input signal at span calibration"-"Load cell input signal at zero calibration"(dCAL02)
・
The calibration weights at span calibration (dCAL03)