Page
53
AD-4410
9.2.4. RS-232C (AD4410-04) (ch2)
Optionally installed on the AD-4410.
Transmission form
: Start-stop synchronous, bi-directional, half duplex
Baud rate
: 600, 1200, 2400, 4800, 9600, 19200, 38400 bps
Data bits
: 7 or 8 bits
Parity
: 1 bit, Even or Odd (Data bits 7 bits)
None (Data bits 8 bits)
Start bit
: 1 bit
Stop bit
: 1 bit
Code :
ASCII
Terminator
: CR LF, OR (CR:0Dh, LF: 0Ah)
Connector
: 7-pin DIN connector
Connection
Pin No. Signal name Direction
Description
1 TXD Output
Transmission
data
2 RXD Input
Received
data
4 SG -
Signal
ground
6 DSR Output
7
Not used (No connection)
3,5
Current loop output
Shell FG
- Shield
* Do not connect to the current loop output.
* Connect the RTS and CTS by a cable side.
C.L./OP internal circuit
Adaptive connector JA-TCP0576 (provided)
Fig.13 Option internal circuitry (AD4410-04)
Shell
FG
RXD
DSR
TXD
SG
NC
Internal circuit of the
AD4410-04 RS-232C