
AD-4410
Page
40
7.9. Load Cell Output Adjustment
Add a resistor as shown below to adjust the load cell output.
Use a resistor with a high resistance value and a low temperature coefficient.
Fig.8 Load cell output adjustment
* Because the zero point of the AD-4410 has a wide adjustable range, as long as there
is no problem with a load cell, correcting an output by the AD-4410 is hardly ever
required.
Before an output correction is carried out, confirm load cells (deformation, wiring
mistakes, contact with anything, or model selection etc.) and connections.
"C Err 2"
(When exceeding in the positive direction)
"C Err 3"
When exceeding in the negative direction)
EXC-
SIG+
SIG-
EXC+
EXC-
SIG+
SIG-
EXC+
Re
sist
or
add
ed
Re
sist
or
ad
de
d