AD-4408A
Page
46
Function No.
Parameter or
Setting range
Description
Default
value
User
setting
Calf13
1
to
3
Condition to judge over when the gross
weight is negative.
A/D negative over or
1
: Gross weight < -999999
2
: Gross weight < -999999,
Gross weight < negative weighing capacity
3
: Gross weight < -19 d
1
Calf14
1
to
2
Condition to judge over when the net weight
is negative.
A/D negative over or
1
: Net weight < -999999
2
: Net weight < -999999,
Net weight < negative weighing capacity
1
Calf15
0
to
1
Select whether or not to clear the zero
value.
0
: Disables.
1
: Enables.
1
Calf16
0
to
1
Select whether or not to perform zero upon
power-ON.
0
: Disables.
1
: Enables.
0
Calf17
-7.00000
to
7.00000
Input voltage from a load cell at zero.
X.XXXXX mV/V
0.00000
Calf18
0.00001
to
9.99999
Input voltage from a load cell at span
X.XXXXX mV/V
2.00000
Calf19
1
to
999999
The value of the calibration weight
corresponding to the span input voltage
20000
Calf20
0
to
1
Select whether or not to output when the
gross weight is negative.
0
: Enables output.
1
: Disables output.
0
Calf21
0
to
1
Select whether or not to restrict
network-related communication.
0
: Disables restriction.
1
: Enables restriction.
0