Page
29
AD-4408A
7.5. Load Cell Output Adjustment
Add a resistor as shown below to adjust the load cell output.
Use a resistor with a high resistance value and a low temperature coefficient.
Fig.7 Load cell output adjustment
"C Err 2"
(When exceeding in the positive direction)
"C Err 3"
When exceeding in the negative direction)
EXC-
SIG+
SIG-
EXC+
EXC-
SIG+
SIG-
EXC+
Re
sist
or
add
ed
Re
sist
or
ad
de
d