8. General Functions
General functions are divided into groups per function and are indicated by the group
name with the function number.
NOTE: General functions determine the AD-4408A performance and all of the
settings are stored in the FRAM.
8.1. Setting Procedure
Step 1 While pressing and holding the ENTER key, press the F key.
fnc
is displayed to indicate that the indicator will enter the general function
mode.
Step 2 Press the ENTER key. The indicator enters the general function mode.
To go back to the weighing mode without entering the general function mode,
press the ESC key.
Step 3 Press the
∧
or
∨
key to select the function group to be set.
Display Group
name
fncf
Basic function
Cl f
Standard serial output
∗∗∗
f
∗∗∗
f
∗∗∗
f
Interfaces
Refer to the instruction manual for each interface,
available on our website.
Press the ENTER key. The function number will be displayed.
Step 4 Press the
∧
or
∨
key to select the function number to be set.
Press the ENTER key. The current setting value will be displayed.
Step 5 Change the setting value using either one of the methods below.
Method Description
Selecting a parameter
Only the parameter number to be selected is
displayed and blinks.
Press the
∧
or
∨
key to select a parameter.
Inputting the value
All the digits are displayed and a digit to be changed
blinks.
Press the
<
or
>
key to select a digit and
press the
∧
or
∨
key to change the value.
After setting, press the ENTER key. The next function number is displayed.
When the parameter is not to be changed, press the ESC key to return to the
function number display.