7.4. Description Of The Item "Data Output Mode"
The parameter setting of "Data output mode (
prt
)" applies to the performance when the data is
transmitted using the RS-232C interface
.
Normal mode
The output is performed only when receiving a command such as a Q command.
Required
setting
dout
prt
0
Normal
mode
Auto print modes A and B
When the displayed value is stable and the conditions of "Auto print polarity", "Auto print
difference" and reference value are met, the balance outputs the weighing data once.
Auto print mode A
Example
For weighing each time a sample is placed and removed, with "
ar-d
"
set to "
1
" (to adjust zero after the data is output).
Required
setting
dout
prt
1
Auto print mode A (reference = zero)
dout
ap-p
Auto print polarity
dout
ap-b
Auto print difference
dout
ar-d
on
Zero after output
Auto print mode B
Example
For weighing while a sample is added.
Required
setting
dout
prt
2
Auto print mode B (reference = last stable value)
dout
ap-p
Auto print polarity
dout
ap-b
Auto print difference
Stream mode
The balance outputs the weighing data continuously regardless of the display condition. When
the display refresh rate is set to 5 times / second (
5pd
0
), the data output rate is also set to the
same 5 times / second. The display does not blink in this mode.
Example
For monitoring data on a computer.
Required
setting
dout
prt
3
Stream
mode
ba5fnc
5pd
Display refresh rate
5if
bp5
Baud
rate
Caution
The balance may not transmit the data completely at the specified refresh rate, depending
on the baud rate or time, date and ID number appended to the weighing data.
Interval mode
The weighing data is periodically output.
Example
In case of monitoring the weighing value periodically, the time and date
can be added to the weighing data depending on settings of
5-td..
Required
setting
dout
prt
3
Interval
mode
dout
int
Interval
time