7. FUNCTION
TABLE
The function table reads or rewrites the parameters that are stored in the balance.
These parameters are maintained in non-volatile memory, even if the AC adapter is removed.
The function table menu consists of two layers. The first layer is the "Class" and the second layer
is the "Item".
7.1. How to set
(
When using the AD-1691
)
Using the provided AD-1691 weighing environment analyzer as a remote controller enables
changes or confirmation of function settings.
1. Connect the RS-232C connector for the A&D's external display to the AD-1691's arbitrary
RS-232C connector using the cable.
2. Connect the AD-4212D to the power using the AC adapter to turn on the power and have the
AD-4212D enter the weighing mode.
3. Connect the AD-1691 to the power using the AC adapter and turn the power on.
4. Perform the balance function settings at the AD-1691.
5. Touch the Take Data button to start communication between the AD-1691 and AD-4212D.
6. The screen shows the function settings list.
7. Click the arrow mark to change the setting values.
* The changed contents have yet to be confirmed at this step.
8. Touch the Confirm button to start the communication and change the balance function settings.
The device will be restarted when the setting changes were ended.