background image

 

 

 

 

Sampling Using the Pump 

 
1)

 

Break off both ends (tips) of the detector tube by using the provided tube tip breaker or the tube 

tip holder (GAS721). In the case of a twin tube, break off both ends of both tubes and connect the 
ends of the tubes marked with a 

©

 using the rubber connector.  

  

  

How to break off the tube tip 

 

 

Insert  the  detector  tube  end  straight  into  the  tube  tip  breaker,  and  rotate  the 
tube for one full turn so that the diamond cuter of the tube tip breaker scratches 
the surface of the tube end. Hold the detector tube firmly at a point close to the 
tube tip breaker and bend the tube toward the handle to break the tube tip.  

 
 
 

2)

 

Confirm  the  pump  handle  is  fully  pushed  in  (therefore  the  guide  line  on  the  pump  shaft  is  not 

seen). Then insert the detector tube into the rubber inlet of the pump with the arrow (     ) on the 
tube pointing toward the pump.  
 

 
 
 
 
 
 
 
 
 
3)

 

Align the guide mark (red line) on the back plate and the guide mark (

100 or 

50) depending 

on the handle.  
NOTE: The guide mark 

100 means a full pump stroke (100ml) and the guide mark 

50 means ½ 

a pump stroke (50ml). You can confirm the number of pump strokes of a given detector tube by 
referring  to  the  guide  marks  on  the  tube  body  or  by  referring  to  the  detector  tubes  operation 
manual (HBE)/Instruction leaflet supplied in a pack of tubes. 
 

 
 

 

 

Summary of Contents for Gastec GV-100S

Page 1: ...ml and the half stroke 50ml positions are marked exactly by the red line on the pump shaft and the handle is precisely locked at those positions If you pump fully n times by allowing for sampling time...

Page 2: ...of the flow finish indicator at this time 1 Confirm that the inlet clamping nut is firmly tightened 2 After confirming that the pump handle is fully in therefore the guide line on the pump shaft is n...

Page 3: ...n of the counter and may cause a malfunction 5 Unlock the handle by turning it more than turn and guide it back gradually Confirm the handle returns to the initial position and the guide line on the p...

Page 4: ...nt close to the tube tip breaker and bend the tube toward the handle to break the tube tip 2 Confirm the pump handle is fully pushed in therefore the guide line on the pump shaft is not seen Then inse...

Page 5: ...Wait until the specified sampling time has elapsed The completion of the sampling of 100ml or 50ml can be confirmed by the flow finish indicator on the handle Unlock the handle by turning it more than...

Page 6: ...counter ring with the white mark Rotate the outer ring of the counter to align the number with the white mark 2 Each time you do a sample the number will increase by one After you have completed samp...

Page 7: ...to the manual of each detector tube and correct for the ambient pressure Also check for any interference gases present in the measuring environment If there are any external influences such as gas tha...

Page 8: ...e 1 Temperature correction using correction factor To find the true concentration when the tube reading is 2 5ppm at 15 C 59 F first using the table below obtain the correction factors for the tempera...

Page 9: ...eading requires humidity correction with the measurement temperature of 0 C to 40 C 32 F to 104 F read the ambient humidity at the sample point using the no 6 Gastec Water Vapour detector tube Example...

Page 10: ...the number of pump strokes n required and if applicable the stroke correction factors see example GAS3L on the next page below Multiply the indicated value by an appropriate factor to obtain the true...

Page 11: ...similar chemical properties as the target gas If such interference gas exists measurement will be incorrect Be sure to check if interference gases are present in the measurement environment and if pr...

Page 12: ...t from inside the cylinder and around the perimeter of the piston with a soft cloth 3 Apply the lubricant evenly on the inside of the wall at the opening of the cylinder 4 Push the piston into the cyl...

Page 13: ...or breaking off tube tips please contact us for a replacement head block Repair of the counter When any abnormality is found with the counter please contact us Never give a gastec precision pump any s...

Reviews: