background image

 

 

 

 

 

Correction for Atmospheric Pressure  

Tube readings are affected by significant fluctuations of atmospheric pressure. All Gastec detector 
tubes are calibrated based on a normal atmospheric pressure of 1013hPa (760mmHg) and their 
indications will not be affected over the range of 

±

 10% of normal pressure (912 to 1114hPa or 684 to 

836mmHg). If the pressure at the time of measurement is not within this range, correct the tube 
reading as follows: 

True concentration 

 

 

 

 

Possible effects from interference gases 

Detector  tubes  use  chemical  reactions  to  indicate  colour  changes.  Therefore,  indications  on  the 
detector tubes may be affected when there is an interference gas that has similar chemical properties 
as  the  target  gas.  If such  interference  gas exists, measurement  will  be  incorrect.  Be  sure  to  check  if 
interference  gases  are  present  in  the  measurement  environment  and  if  present,  confirm  before 
measurement  with  the  operation  manual  of  the  detector  tube  and  its  possible  effects  on  the 
measurement.  

The table  on  the  leaflet  included in each pack  detector tubes  shows  interference  properties  of each 
gas  monitored  (if  applicable).  Some  materials of  specific  concentration  are  not  shown on  the  leaflet 
may affect the reading under specific conditions.  

Maintenance

 

Frequent use of the Gastec precision pump can cause an air leak which can be easily resolved. Most 
common problems are loose inlet clamping nut, damaged or deteriorated rubber inlet, and 
deteriorated or insufficient lubricant. The table below shows the possible causes and solutions: 

 

 

 

Summary of Contents for Gastec GV-100S

Page 1: ...ml and the half stroke 50ml positions are marked exactly by the red line on the pump shaft and the handle is precisely locked at those positions If you pump fully n times by allowing for sampling time...

Page 2: ...of the flow finish indicator at this time 1 Confirm that the inlet clamping nut is firmly tightened 2 After confirming that the pump handle is fully in therefore the guide line on the pump shaft is n...

Page 3: ...n of the counter and may cause a malfunction 5 Unlock the handle by turning it more than turn and guide it back gradually Confirm the handle returns to the initial position and the guide line on the p...

Page 4: ...nt close to the tube tip breaker and bend the tube toward the handle to break the tube tip 2 Confirm the pump handle is fully pushed in therefore the guide line on the pump shaft is not seen Then inse...

Page 5: ...Wait until the specified sampling time has elapsed The completion of the sampling of 100ml or 50ml can be confirmed by the flow finish indicator on the handle Unlock the handle by turning it more than...

Page 6: ...counter ring with the white mark Rotate the outer ring of the counter to align the number with the white mark 2 Each time you do a sample the number will increase by one After you have completed samp...

Page 7: ...to the manual of each detector tube and correct for the ambient pressure Also check for any interference gases present in the measuring environment If there are any external influences such as gas tha...

Page 8: ...e 1 Temperature correction using correction factor To find the true concentration when the tube reading is 2 5ppm at 15 C 59 F first using the table below obtain the correction factors for the tempera...

Page 9: ...eading requires humidity correction with the measurement temperature of 0 C to 40 C 32 F to 104 F read the ambient humidity at the sample point using the no 6 Gastec Water Vapour detector tube Example...

Page 10: ...the number of pump strokes n required and if applicable the stroke correction factors see example GAS3L on the next page below Multiply the indicated value by an appropriate factor to obtain the true...

Page 11: ...similar chemical properties as the target gas If such interference gas exists measurement will be incorrect Be sure to check if interference gases are present in the measurement environment and if pr...

Page 12: ...t from inside the cylinder and around the perimeter of the piston with a soft cloth 3 Apply the lubricant evenly on the inside of the wall at the opening of the cylinder 4 Push the piston into the cyl...

Page 13: ...or breaking off tube tips please contact us for a replacement head block Repair of the counter When any abnormality is found with the counter please contact us Never give a gastec precision pump any s...

Reviews: