39
Specifica
tions
/
38
Specifica
tions
/
/ 14
Specifications
Product specification
Measurement range
1.1–33.3 mmol/L
Sample size
Minimum 0.4 μL
Test time
5 seconds
Sample type
•
Fresh capillary whole blood.
•
Fresh venous whole blood
(healthcare professionals only).
Calibration
Plasma-equivalent
Assay method
Electrochemical
Battery life
1,000 tests
Power
Two 3.0 V lithium batteries (disposable, type
CR2032)
Memory
1,000 test results
Size
103 x 54 x 15.9 (mm)
Weight
71.5 g (with batteries)
Bluetooth
®
technology
•
Frequency range: 2.4–2.4835 GHz
•
Operating range distance: maximum 10 me-
ters (unobstructed)
•
Operating channels: 40 channels
•
Security encryption: 128-bit AES
(Advanced encryption standard)
Operating ranges
Temperature
5–45 ˚C (41–113 ˚F)
Relative humidity
10–90 %
Hematocrit
15–65 %
Storage conditions
Glucose meter
(with battery)
0–50 ˚C (32–122 ˚F)
Test strip
1–30 ˚C (34–86 ˚F)
The GlucoMen Day METER is in compliance with Directive 2014/53/EU
on radio equipment. The full text of the EU Declaration of Conformity is
available at the following internet address:
www.red.menarinidiagnostics.com