UFC-420/440 water Filtration system

9

P/N 1010908 Rev. A 03/12

A.J. Antunes & Co.

The unit also must be flushed to remove air and the 

shipping/storage solution.  For maximum quality, the 

permeate water produced during the flushing procedure 

must be discarded.  Direct this permeate water to drain.  

  1.  Plug the power cord into the appropriate electri-

cal outlet.  The display will power on and the LED 

display will display the following for about two sec-

onds each:

 

• 

8.8

  

• 

F8

 followed by its time settings in seconds 

• 

Fi

 and its time setting in minutes or hours.

 The 

F8

 and 

Fi

 sequence repeats for 30 seconds 

after which the unit automatically returns to the 

Flush Interval Mode (

Fi

).  The time setting for 

Fi

 will 

be displayed and the decimal point will flash in one 

second intervals.

  2.  open the tap or faucet closest downstream to the 

filter system.

  3.  Slowly open the inlet water valve and allow water 

to enter the system.

  4.  Press and hold the Start button.  After 6-7 sec-

onds, the drain valve will open and 

FL

 will appear 

on the display.

  5.  Continue holding the Start button down for at least 

30 seconds to keep the drain open.  This flushes 

air out of the center of the hollow fibers.  The 

drain will remain open as long as the Start button 

is pressed.  Check to make sure that the drain 

water is directed into the drain without splashing.

  6.  Release the Start button.  This closes the drain 

valve.  Water should continue to flow through the 

system and out of the open tap.  Allow water to 

flow out of the tap for at least 15 minutes at maxi-

mum flow rate.

  7.  Close the tap and let the system stand with no 

water flow for 15 minutes to allow any trapped air 

to come out of the hollow fibers.  Check for leaks 

at all fittings.

caUtIOn

Ingesting the protective solution may cause irrita-

tion of the gastrointestinal tract, colic, diarrhea, or 

other similar symptoms

InstallatIOn (continued)

  8.  After 15 minutes without water flow, open the 

tap for 5 minutes to allow any trapped air to be 

flushed out.

  9.  Close the tap.  Flushing is complete.

sanItIZInG tHe sYstem anD lInes

The plumbing must be sanitized to eliminate possible 

contamination that may have occurred during the instal-

lation process.  Chlorine bleach can be used to sanitize 

the plumbing.  The amount of bleach to use depends 

on the system installed and the amount of plumbing 

downstream of the filter system.  Generally, 1 or 2 

ounces (30 ml) of bleach will be sufficient to sanitize the 

system.

  1.  Make sure that the system has been flushed of 

air and debris as described in the Flushing and 

Starting up the System section of this manual.

  2.  open the tap closest downstream to the filter sys-

tem.

  3.  Close the inlet water valve and allow the system 

to depressurize.

  4.  Place a bucket under the strainer at the inlet 

connection to the system.  open the strainer by 

unscrewing the cap.  Water will flow out of the 

strainer as the system drains.

  5.  When the water flow out of the strainer stops, 

pour the liquid bleach into the strainer.  Be careful 

not to spill bleach onto clothing or skin.  You may 

want to add the bleach using a cup.  Reattach the 

cap on the strainer.

  6.  Slowly open the inlet water valve and allow water 

to flow out of the tap until the smell of bleach is 

present.

  7.  Close the tap and let the system stand with no 

water flow for at least 15 minutes to allow the 

bleach to sanitize the pipes.

  8.  After 15 minutes without flow, open the tap and 

flush until the presence of bleach is gone.  All 

other taps should be opened to flush any bleach 

from the plumbing.

  9.  Close the tap.  The sanitization is complete.

Program the timer following the procedure outlined in 

the operation section of this manual.

Summary of Contents for 9700610

Page 1: ...Water filtration system UFC 420 440 Series Owner s Manual P N 1010908 Rev A 03 12 Manufacturing Numbers 9700610 9700612 9700614 9700616 ...

Page 2: ...t arrives damaged contact the carrier imme diately and file a damage claim with them Save all packing materials when filing a claim Freight damage claims are the responsibility of the purchaser and are not covered under warranty The warranty does not extend to Damages caused in shipment or damage as result of improper use Installation of electrical service Normal maintenance as outlined in this ma...

Page 3: ...the equipment is furnished with a properly grounded cord connector Do not attempt to defeat the grounded connector Install or locate the equipment only for its intend ed use as described in this manual Do not use corrosive chemicals in this equipment Do not operate this equipment if it has a dam aged cord or plug if it is not working properly or if it has been damaged or dropped This equipment sho...

Page 4: ...d person This equipment is to be installed to comply with the local plumbing code and any other applicable code Water pressure must not exceed the membrane differential pressure of 45 psig 3 1 bar To reduce water pressure install a water pressure regulator and set the system inlet water pressure to 45 psi 3 1 bar Inlet Water Pressure psi Inlet Turbidity NTU Required Additional Equipment Installed ...

Page 5: ...this appliance Check with a qualified electrician if you are in doubt as to whether the appliance is properly grounded Specifications Filter Cartridges Maximum Operating Pressure 3 bar 45 psi Maximum Operating Temp 40 C 104 F Trans Membrane Pressure 0 5 2 5 Bar 7 36 psi pH Range 3 10 MWCO 100 kD Sanitizing Temp 80 C 176 F Unit A Width B Depth C Height Operating Weight w water UFC 420 10 inches 254...

Page 6: ...UFC 420 440 water Filtration system 6 P N 1010908 Rev A 03 12 A J Antunes Co Specifications Figure 2 Components Components LED Display Program Button Power Cord Water Inlet Drain Permeate Product Water ...

Page 7: ...NOTE This unit is designed to use tap water not to exceed 104 F 40 C The UFC 420 and UFC 440 systems use the following connections Figure 1 Water Inlet 3 4 NPT Permeate Product Water 3 4 NPT Drain 3 4 NPT Installation When making a plumbing connection to the system remember to use a back up wrench on the supporting plumbing Always use a good quality approved pipe sealant or thread seal tape on pip...

Page 8: ...stem hold a bucket or other con tainer at the inlet water line and slowly open the inlet water valved Allow the pipe to flush until all debris is removed Permeate Line Plumbing To ensure the highest quality and safest water it is recommended that a check valve to prevent backflow be installed in the water line after the permeate connec tion This will help prevent possible contamination of the filt...

Page 9: ...gesting the protective solution may cause irrita tion of the gastrointestinal tract colic diarrhea or other similar symptoms Installation continued 8 After 15 minutes without water flow open the tap for 5 minutes to allow any trapped air to be flushed out 9 Close the tap Flushing is complete Sanitizing the System and Lines The plumbing must be sanitized to eliminate possible contamination that may...

Page 10: ...order to keep water filtering during a power outage Figure 4 LED Display Front Panel LED Display Fi Flush Interval Mode is the time between flushings and is displayed in minutes or hours Fd Flush Duration Mode is the amount of time used to flush and remove debris from filter cartridge and is displayed in seconds FL Manual Flush Mode operation solenoid is manually activated by the user Starting the...

Page 11: ...returns to Flush Interval Mode NOTE Fi times are displayed in ten minute inter vals after fifty minutes they are displayed in hours Table A Figure 7 LED Display for Flush Duration Mode Figure 8 Time Setting Fd Mode 55 Seconds Table A LED Display Settings Figure 6 Time Setting Fi Mode 22 Hours Figure 5 LED Display for Flush Interval Mode Setting Flush Duration FD 2 When Fd is displayed press and re...

Page 12: ...e intervals after fifty minutes they are displayed in hours Table B TO SET FD FLUSH DURATION MODE 1 Press and hold down start button for four seconds Release button LED will display Fd 2 When Fd is displayed Figure 9 press and release button to view current setting 3 To make a change press and hold down start button to scroll through settings release button at desired setting Figure 10 4 After ten...

Page 13: ...t button down for 6 7 seconds until the LED displays FL Figure 11 Continue holding the button down to keep the drain open Drain will remain open for as long as start button is depressed FL will flash on LED Releasing the but ton closes the drain and puts the unit back into startup mode After ten seconds if the start button is not pressed the unit automatically returns to the Flush Interval Mode Fi...

Page 14: ...en the strainer by unscrewing the cap Water will flow out of the strainer as the system drains 4 Press and hold the Start button to help drain the system 5 Reattach the cap on the strainer when the water flow stops Unplug the power cord 6 Using a strap wrench up to 6 capacity loosen the top and bottom end caps DO NOT remove the top and bottom End Caps yet 7 Disconnect the permeate port connection ...

Page 15: ...al growth in pipes Sanitize plumbing Water conditions changed Consider installing taste and odor filtration Broken capillary in Filter Cartridge Replace Filter Cartridge Flush runs continuously Drain Valve stuck open Replace rebuild the Drain Valve Controller sending continuous signal to valve Replace the controller Flush runs too long Program duration set too long Re program the unit to flush for...

Page 16: ...UFC 420 440 water Filtration system 16 P N 1010908 Rev A 03 12 A J Antunes Co Replacement parts UFC 420 440 10 9 11 16 17 16 18 15 26 27 1 28 5 2 12 22 23 4 3 24 25 19 20 UFC 420 440 ...

Page 17: ...3 Mounting Bracket Weldment UFC 420 1 0021332 Mounting Bracket Weldment UFC 440 1 12 0021455 Main Housing Weldment UFC 420 1 0021460 Main Housing Weldment UFC 440 1 15 2090122 Cartridge Filter 4 x 20 1 2090123 Cartridge Filter 4 x 40 1 16 0011606 End Cap Union Connector Assy 4 2 17 2190127 Union End Cap 4 2 18 7000405 O Ring End Cap Union 2 19 2190123 Y Strainer 3 4 NPT 1 20 2090126 Screen Y Strai...

Page 18: ...0908 Rev A 03 12 A J Antunes Co Wiring Diagram TIMER BOARD SOLENOID VALVE RED BLK T2 T3 T4 WHT BLK GND TERMINAL BLOCK GRN YEL WHT BLU BLK BRN POWER CORD NOTE ALL WIRED TO BE 16 GA AWM 105ºC UNLESS OTHERWISE SPECIFIED 14 GA AWM 105ºC 22 GA AWM 105ºC ...

Page 19: ...UFC 420 440 water filtration system 19 P N 1010908 Rev A 03 12 A J Antunes Co Notes ...

Page 20: ... operating and main tenance instructions of this manual and or labor involved in moving adjacent objects to gain access to the Equipment 2 A J Antunes Co reserves the right to make changes in design or add any improvements on any product The right is always reserved to modify equipment because of factors beyond our control and government regulations Changes to update equipment do not constitute a ...

Reviews: