Safety First    

 

The  PhotoFluor  II  family  of  light  sources  use  a  powerful  lamp  that 
produces  heat  as  well  as  visible  and  UV  light.  Proper  care  must  be 
taken in the setup and operation to prevent injury.  

 

 
 

The lamp should only be handled wearing gloves 
and suitable eye protection.

   

 

 

 

When  operating,  the  lamp  produces  significant 
amounts  of  ultraviolet  radiation.    Never  look 
directly  at the lamp or at light reflected from the 
lamp  or  the  output  from  the  liquid  light  guide. 

 

Do  not  expose  skin  directly  to  the  output  of  the 
liquid light guide.

 

 
 

WARNING:  

Use  only  the  recommended  bulb  replacement.  
Never 

substitute 

the 

bulb 

with 

another 

manufacturer’s bulb. 

 
 

WARNING:  

DON’T  BE  A  FOOL,  LET  IT  COOL. 

The  lamp 

produces  significant  amounts  of  heat.  Always  allow 
the  light  source  components  to  cool  to  ambient 
temperature  before  attempting  any  adjustments  or 
replacement of parts. 

 

WARNING:   

After 1500 hours of operation, the metal halide lamp 
must  be  replaced  to  ensure  suitable  performance 
and  safety.    The  light  source  will  alert  the  user  at 
1500 hours that the lamp is due for replacement.   
At 1600 hours, the engine will cease operation as 
the bulb is no longer safe to use. 

 

WARNING:   

Used  lamps  should  not  be  discarded  in  the 
conventional  waste  stream.    The  lamps  contain 
metals that must be handled as potentially toxic.  

 
 

 

 

Please consult the following web address for current 
information:  

www.lamprecycle.org

. 

 

 

 

 

Summary of Contents for PhotoFluor II

Page 1: ...r II 10999 000 000 and PhotoFluor II NIR 10998 000 000 light sources are designed and tested in accordance with ANSI UL 61010 1 CAN CSA C22 2 Number 61010 1 CENELEC EN 61326 1 2006 and FCC Part 15 Subpart B Class B 89 North customer service or your authorized 89 North distributor should be informed immediately in the event of any damage or malfunction in the equipment 89 North 1 Mill St Unit 285 B...

Page 2: ...Never substitute the bulb with another manufacturer s bulb WARNING DON T BE A FOOL LET IT COOL The lamp produces significant amounts of heat Always allow the light source components to cool to ambient temperature before attempting any adjustments or replacement of parts WARNING After 1500 hours of operation the metal halide lamp must be replaced to ensure suitable performance and safety The light ...

Page 3: ...he bulb except the ceramic base and power wires as improper handling may shorten bulb life and may lead to catastrophic failure of the lamp WARNING LAMP EXPLOSION HAZARD The lamp contains toxic gasses under high pressure In the event of an explosion mercury vapor high temperature gasses and very fine glass may exit the unit All persons should leave the surrounding area at once to prevent mercury i...

Page 4: ...he lamp installed When removing the lamp take care not to touch the reflector surface inside or outside as finger oils or contaminants will harm the bulb The lamp s connector must be properly installed After installing the bulb and the unit is turned on you must reset the lamp counter The bulb is pictured below The lamp should only be handled wearing gloves and suitable eye protection Reflector La...

Page 5: ...iver and pulling the cover up and off Next remove the inner lamp chamber cover by removing the thumbscrew and sliding the plate off See Figure 2 STEP 3 DISCONNECT the lamp power cable brown connector Next locate the lamp mount and disengage the metal retaining wire from the tabs See Figures 3 4 Remove the lamp and dispose of properly Thumbscrew Figure 2 Lamp chamber cover Lamp chamber cover slides...

Page 6: ...the upright 12 o clock position See Figure 3 Try to move the lamp side to side to assure it is properly seated in the recessed circular opening STEP 5 ENGAGE the metal retaining wire and place them within the tabs The spring action wire will secure the lamp in place STEP 6 CONNECT the lamps power cable brown connector See Figures 3 4 Figure 3 Top view of lamp chamber Lamp power connector Lamp powe...

Page 7: ...ault Select response using left and right arrow keys and press the green center Enter button If YES is selected a message will appear on the display asking if you are sure YES or NO flashing NO is default Select response using left and right arrow keys and press the green center Enter button The lamp hour counter is now reset NOTE The lamp hour reset menu can only be accessed during the 3 minute l...

Page 8: ...ribed in this manual The lamps have an expected life span of at least 1500 hours and will be replaced if they fail or provide poor performance during that time provided they are handled as described herein The liquid light guide is under warranty for one year from delivery Contact Information 89 North 1 Mill St Burlington VT 05401 U S A 877 417 8313 1 802 881 0302 www 89north com ...

Reviews: