background image

 

 

 

- 7 - 

- 7 -7 

Rev B

 

7

 

 

 

 
 
 

STEP 7

  

REATTACH

  the  lamp  chamber  cover  with  the 

thumbscrew.  Carefully  reposition  the  outer  cover 
and reattach the four screws. 

 
 

STEP 8

 

PLUG  THE  UNIT  IN  AND

 

TURN  ON

  the  power 

switch. Reset the lamp hour counter by pressing and 
holding  the  up  arrow  button  during  warm-up.  A 
message  appears  on  the  display  asking  if  you  want 
to reset the counter

; “

YES

’ or “

NO

” (flashing “NO” is 

default).  Select  response  using  left  and  right  arrow 
keys  and  press  the  green  center 

“Enter”  button.  If 

YES

”  is  selected,  a  message  will  appear  on  the 

display  asking  if  you  are  sure

;  “

YES

’  or  “

NO

” 

(flashing “NO” is default). Select response using left 
and  right  arrow  keys  and  press  the  green  center 
“Enter” button. The lamp hour counter is now reset. 

 

       NOTE: 

The lamp hour reset menu can only be 
accessed during the 3 minute lamp warm-up 
period.  Pressing and holding the up arrow 
button in normal operating mode will not initiate 
the lamp hour reset sequence. 

 

Lamp power 
connector 

Figure 4: Lamp connector 

Summary of Contents for PhotoFluor II

Page 1: ...r II 10999 000 000 and PhotoFluor II NIR 10998 000 000 light sources are designed and tested in accordance with ANSI UL 61010 1 CAN CSA C22 2 Number 61010 1 CENELEC EN 61326 1 2006 and FCC Part 15 Subpart B Class B 89 North customer service or your authorized 89 North distributor should be informed immediately in the event of any damage or malfunction in the equipment 89 North 1 Mill St Unit 285 B...

Page 2: ...Never substitute the bulb with another manufacturer s bulb WARNING DON T BE A FOOL LET IT COOL The lamp produces significant amounts of heat Always allow the light source components to cool to ambient temperature before attempting any adjustments or replacement of parts WARNING After 1500 hours of operation the metal halide lamp must be replaced to ensure suitable performance and safety The light ...

Page 3: ...he bulb except the ceramic base and power wires as improper handling may shorten bulb life and may lead to catastrophic failure of the lamp WARNING LAMP EXPLOSION HAZARD The lamp contains toxic gasses under high pressure In the event of an explosion mercury vapor high temperature gasses and very fine glass may exit the unit All persons should leave the surrounding area at once to prevent mercury i...

Page 4: ...he lamp installed When removing the lamp take care not to touch the reflector surface inside or outside as finger oils or contaminants will harm the bulb The lamp s connector must be properly installed After installing the bulb and the unit is turned on you must reset the lamp counter The bulb is pictured below The lamp should only be handled wearing gloves and suitable eye protection Reflector La...

Page 5: ...iver and pulling the cover up and off Next remove the inner lamp chamber cover by removing the thumbscrew and sliding the plate off See Figure 2 STEP 3 DISCONNECT the lamp power cable brown connector Next locate the lamp mount and disengage the metal retaining wire from the tabs See Figures 3 4 Remove the lamp and dispose of properly Thumbscrew Figure 2 Lamp chamber cover Lamp chamber cover slides...

Page 6: ...the upright 12 o clock position See Figure 3 Try to move the lamp side to side to assure it is properly seated in the recessed circular opening STEP 5 ENGAGE the metal retaining wire and place them within the tabs The spring action wire will secure the lamp in place STEP 6 CONNECT the lamps power cable brown connector See Figures 3 4 Figure 3 Top view of lamp chamber Lamp power connector Lamp powe...

Page 7: ...ault Select response using left and right arrow keys and press the green center Enter button If YES is selected a message will appear on the display asking if you are sure YES or NO flashing NO is default Select response using left and right arrow keys and press the green center Enter button The lamp hour counter is now reset NOTE The lamp hour reset menu can only be accessed during the 3 minute l...

Page 8: ...ribed in this manual The lamps have an expected life span of at least 1500 hours and will be replaced if they fail or provide poor performance during that time provided they are handled as described herein The liquid light guide is under warranty for one year from delivery Contact Information 89 North 1 Mill St Burlington VT 05401 U S A 877 417 8313 1 802 881 0302 www 89north com ...