Page 12
3.5 Left-to-Right Adjustment
1. Remove the cover.
2. Locate the lower M4 screw underneath the platen that receives the PCR / microplate (see
Figure 10). If the unit is under-sealing on the right, loosen the screw and tilt the lift to the left.
If the unit is under-sealing on the left, loosen the screw and tilt the lift to the right.
3. Re-tighten the screw.
Figure 12 - Left-to-Right Adjustment