4.10 TRIOS System Environment

IMPORTANT NOTICE
Operating Conditions

Ambient temperature range: +15 – +30 °C 
Relative humidity: 10 – 85 % (non-condensing) 
Atmospheric pressure: 800 – 1100 hPa
The scanner should be used in an environment  
maintaining a consistent room temperature .

IMPORTANT NOTICE
Storage Conditions

Ambient temperature range: -10 – +60 °C 
Relative humidity: 10 – 85 % (non-condensing)  
Atmospheric pressure: 800 – 1100 hPa

IMPORTANT NOTICE
Transport Conditions

Ambient temperature range: -10 – +60 °C 
Relative humidity: 10 – 85 % (Non-condensing) 
Atmospheric pressure: 620 – 1200 hPa

4.11 Technical specifications

Light source 
specifications

Do not view directly with optical instruments, such 
as spectacle microscopes . The product complies 
with EN (IEC) 62471 Photobiological safety of lamps 
and lamp systems .

Sterilization 
and high-level 
disinfection 
method

Scanner tips can be autoclaved  
or high-level disinfected .

Power input

C4.0/ C4.1 MOVE/MOVE+ cart (for S3A)

Power input: 100-240 V AC, 50/60 Hz,  
180 VA max Insulation class: class I

PS3-2 pod (for wired S3P)

External medical-grade power supply, rated 30 
W max Power input: 100-240 V, 1 .0 A, 50/60 Hz 
Insulation class: Class I

C3BP: Battery charger 

Power input: 100-240 V, ~0 .3-0 .7 A, 50/60 Hz  
Protection class: IP20 
Insulation class: Class II

S3-2 scanners

 .

 

Wireless- battery powered 

S3-2 scanners

 . Wired- TRIOS 4 cord

3Shape TB-02 Rechargeable Battery

  

3,6 V, 2000 mAh, 7 Wh

3Shape TB-03

 

Rechargeable Battery

  

3 .6 V, 3450 mAh, 12 Wh

Wireless 
Module

The scanner contains an IEEE 802 .11ac module,  
FCC ID: PD98265NG

Wireless 
Transmit Power

Maximum 100 mW

Wireless 
Frequency Band

2 .401-2 .483 GHz

5 .15-5 .725 GHz  
(the actual frequencies are dependent on local reg-
ulations and the configuration of the product .)

5. Safety and Warnings

5.1 Prerequisites

CAUTION

Read all instructions carefully, including all warnings  
and caution notices . You must comply with the warnings  
in this manual to prevent injury to persons and damage  
to equipment . Proper functionality and safety can only  
be guaranteed if the safety precautions in this safety guide 
and on the TRIOS IOS system are observed .

Country dependent feature:

 Fluorescence scanning mode .

CAUTION

In fluorescence scanning mode:
•  Limit exposure of the oral mucosa .
•  Limit scanning time to 15 seconds per tooth .
•  Take appropriate measures for eye protection .
•  Only run the scanner in fluorescence mode while  

it is in the patient’s mouth .

•  Do not direct the blue light emitted by the scanner 

towards the eyes of the patient or operator .

•  Consider using eyewear that blocks blue light .
•  Consider viewing the computer screen while scanning 

instead of the mouth of the patient .

Preventive inspection before use of the system

CAUTION

Please examine the TRIOS IOS system for any mechanical 
damage on:
•  all enclosures,
•  all cables .
Safety can only be guaranteed if NO DAMAGE  
on the TRIOS IOS system is observed .

Modifications of the system

CAUTION

Modifications of the system which may affect the safety of 
the operator, patients or third parties are prohibited by law!

Approved software only

CAUTION

Install only approved software – to prevent interference 
with the runtime reliability of the system and programs 
within it .

Proper training

WARNING

Before you attempt to use the TRIOS scanner equipment 
with patients:
•  You must be trained to use the system and have read and 

understood all sections of this guide describing correct 
operation .

•  You must also be thoroughly familiar with the safe opera-

tion of the TRIOS scanner equipment as described in this 
documentation .

Summary of Contents for TRIOS S3P-2

Page 1: ...04 2021 3Shape TRIOS Safety and Setup Guide S3P 2 S3P 8 S3A 10 S3A 14 S3A 18 S3A 22...

Page 2: ...rmity EU 3ShapeTRIOSA SherebydeclaresthattheTRIOSobservestheprovisionsof Medical Device Regulation EU 2017 745 MDR Restriction of Hazardous Substances in Electrical and Electronic Equipment Directive...

Page 3: ...Electronic Equipment Directive 2012 19 EU Radio Equipment Directive 2014 53 EU CAUTION EU only CE marking for connected products Further products which are connected to this unit must also bear the CE...

Page 4: ...t in PC 3 scanner batteries and a scanner battery charger You can find the figures in the final section of this guide All system variants comprise Handheld TRIOS intraoral scanner S3 2 S1 and S2 Wired...

Page 5: ...our service provider if there is visible damage or malfunctioning System S3A MOVE MOVE cart only Please follow the steps below for assembling TRIOS IOS system Figure 5 Step 1 Place the MOVE MOVE cart...

Page 6: ...is due Other messages include a reminder to mount the scanning tip Please make sure the scanning tip is mounted on the scanner Calibration A calibration of the handheld TRIOS scanner is required befo...

Page 7: ...pendent on local reg ulations and the configuration of the product 5 Safety and Warnings 5 1 Prerequisites CAUTION Read all instructions carefully including all warnings and caution notices You must c...

Page 8: ...it should immediately be calibrated before further use If calibration fails please contact your technical service provider See this guide s instructions on calibrating the handheld scanner WARNING If...

Page 9: ...wer cable network cable and the smart cord must never be subjected to pulling stress Spilled liquids WARNING Do not bring liquids such as beverages near the TRIOS IOS system Do not spill liquids on th...

Page 10: ...S3P pod only WARNING Before each patient make sure any PC connected to the system and its peripherals are disinfected It is recom mended to use medical grade peripherals e g keyboards and mice that ca...

Page 11: ...manual for more information CAUTION New tips must be cleaned and sterilized high level disinfected before using them for the first time CAUTION A class B Vacuum autoclave according to EN 13060 is requ...

Page 12: ...iod of use IMPORTANT NOTICE All surfaces that the protection tip comes in contact with particularly the scanner tube and its optical window must always be cleaned and disinfected before attaching the...

Page 13: ...the correct rating and type 250 V T 3 15 A H must be used To replace the fuses first mount them in the drawer then insert the drawer back into the slot and push all the way in until the tabs click ba...

Page 14: ...3Shape TRIOS A S Holmens Kanal 7 1060 Copenhagen Denmark 10030970A April 2021 Copyright 3Shape TRIOS A S Let s change dentistry together More information at 3Shape com...