WARNING

Use of the UNIT adjacent to or stacked with other equip-
ment should be avoided because it could result in improper 
operation . If such use is necessary, the UNIT and the other 
equipment should be observed to ensure normal operation 
in the used configuration .

WARNING

The product needs special precautions regarding electro-
magnetic compatibility (EMC) and needs to be installed and  
put into service according to the EMC information provided  
in the TRIOS Technical Data .

IMPORTANT NOTICE 

The system is to be used in a Professional Healthcare 
Environment .

4.2 Intended Use

The 3Shape TRIOS 4 IOS system is intended to obtain 3D digital images  
of the topographical characteristics of teeth, including bite, gingivae and 
palate directly from the patient’s mouth or models of the teeth .

CAUTION

Unintended use of the TRIOS IOS system can result in phys-
ical injury to patients, operators and damage the system .

4.3 Users of TRIOS

The 3Shape TRIOS intraoral scanner system is to be operated by legally quali-
fied healthcare professionals in dental clinics . Using the TRIOS IOS system for 
any purpose other than the intended use described above may damage the 
TRIOS IOS system and harm operators and patients . Proper use of the TRIOS 
IOS system includes following the operation and maintenance instructions, 
and observing warnings and caution descriptions in this user guide .

WARNING 
Carefully follow the instructions in this guide

If the instructions for operating the TRIOS IOS system  
as described in this guide are NOT observed, the intended 
protection and safety of the user and patient may be impaired .

4.4 Contra-indications

None

4.5 Patient Target Group

No restrictions in patient population

4.6 System Description

The system is available in these variants:

  

Regional variation can occur regarding the contents of the systems .

S3P-2

Pen grip scanner, supplied with a pod (holder),  
3 scanner batteries, and a scanner battery charger

S3P-8

Pen grip scanner, supplied with a pod (holder)  
that is connected to an external PC via an Ethernet cable, 
and a TRIOS 4 cord

S3A-10

Pen grip scanner, supplied with a MOVE® cart with built-in PC,  
and a TRIOS 4 cord

S3A-14

Pen grip scanner, supplied with a MOVE cart with built-in 
PC, 3 scanner batteries, and a scanner battery charger

S3A-18

Pen grip scanner, supplied with a MOVE®+ cart with 
built-in PC, and a TRIOS 4 cord

S3A-22

Pen grip scanner, supplied with a MOVE+ cart with built-in 
PC, 3 scanner batteries, and a scanner battery charger

You can find the figures in the final section of this guide .  
All system variants comprise:
• 

Handheld TRIOS intraoral scanner, S3-2

 (S1 and S2):  

Wired and wireless handheld dental intraoral scanner  
for scanning the surface of the patient’s teeth . 

• 

TRIOS 4 SmartTip, TST-01

 (S3): Re-usable scanner tip attached to the 

handheld TRIOS Scanner . The scanner tip must be cleaned and sterilized 
prior to first use on a patient and after each use on a patient .

The scanner tip as well as the protection tip (S4) can be attached  
by the following procedure: 
1 . Place the tip over the front tube with the mirror facing upwards  

or downwards . 

2 . Twist the tip until a firm mechanical click is felt .
3 . The tip is removed doing the opposite procedure:
4 . Twist the tip away from the center of the scanner .
5 . Pull the tip away from the scanner body (S4) .

The label on the protection tip indicates the attaching/removing movement 
(S4) . When the tip is detached, the scanner’s optical window (S3) is visible .

The TRIOS 4 tip contains a mirror and electronics . A set of scanner tips  
is delivered with the scanner . The scanner tip is the only applied part .
• 

Protection Tip, TST-03 

(S4): Protects the optical window of the scanner 

from damage and dirt when the scanner is not in use .

• 

Color Calibration Kit, TST-02 

(S5): Used for calibrating TRIOS scanners . 

The kit consists of the color calibration target (S6), the sleeve (S7), and 
the color calibration adaptor (S8) .

• 

TRIOS software:

 Scan application/user interface for the scanner .

• 

Power (mains) cable and plug: 

The system is delivered with a cable 

appropriate for your location . In the USA and Canada, this is a Nema 5-15P 
Hospital Grade mains power cord with plug for the MOVE/MOVE+ cart . 
For the charger a C7 connector is used .

• 

A set of batteries for the scanner, TB-02 or TB-03

 (S9)

• 

A battery charger, C3BP, with wall mount

 (S10)

For systems S3A-10, S3P-8:

• 

TRIOS 4 cord, TSC-01

 (S11): The wired systems are delivered with a cord, 

which is inserted in the scanner by pressing the button as shown in (S2) . 
The blue light indicates a connection to either the Pod or the MOVE+ cart .  
The green light indicates a connection to the scanner .

System S3A furthermore comprises:

• 

MOVE cart C 4.0 and MOVE+ cart C 4.1 (Figure 2), 

for intraoral digital 

impression system .

Following description applies to both MOVE and MOVE+ cart:
•  Has an adjustable monitor (S12, S13, S14) .
•  Has a 

scanner mount (S15) 

for holding the scanner while not in use . 

•  Has 

wheels

 for mobility (and brakes (S16) if mobility is not desired) .

• 

Computer Unit

 . Attached to the MOVE/MOVE+ cart for processing, 

storage and running the TRIOS software .

Summary of Contents for TRIOS S3P-2

Page 1: ...04 2021 3Shape TRIOS Safety and Setup Guide S3P 2 S3P 8 S3A 10 S3A 14 S3A 18 S3A 22...

Page 2: ...rmity EU 3ShapeTRIOSA SherebydeclaresthattheTRIOSobservestheprovisionsof Medical Device Regulation EU 2017 745 MDR Restriction of Hazardous Substances in Electrical and Electronic Equipment Directive...

Page 3: ...Electronic Equipment Directive 2012 19 EU Radio Equipment Directive 2014 53 EU CAUTION EU only CE marking for connected products Further products which are connected to this unit must also bear the CE...

Page 4: ...t in PC 3 scanner batteries and a scanner battery charger You can find the figures in the final section of this guide All system variants comprise Handheld TRIOS intraoral scanner S3 2 S1 and S2 Wired...

Page 5: ...our service provider if there is visible damage or malfunctioning System S3A MOVE MOVE cart only Please follow the steps below for assembling TRIOS IOS system Figure 5 Step 1 Place the MOVE MOVE cart...

Page 6: ...is due Other messages include a reminder to mount the scanning tip Please make sure the scanning tip is mounted on the scanner Calibration A calibration of the handheld TRIOS scanner is required befo...

Page 7: ...pendent on local reg ulations and the configuration of the product 5 Safety and Warnings 5 1 Prerequisites CAUTION Read all instructions carefully including all warnings and caution notices You must c...

Page 8: ...it should immediately be calibrated before further use If calibration fails please contact your technical service provider See this guide s instructions on calibrating the handheld scanner WARNING If...

Page 9: ...wer cable network cable and the smart cord must never be subjected to pulling stress Spilled liquids WARNING Do not bring liquids such as beverages near the TRIOS IOS system Do not spill liquids on th...

Page 10: ...S3P pod only WARNING Before each patient make sure any PC connected to the system and its peripherals are disinfected It is recom mended to use medical grade peripherals e g keyboards and mice that ca...

Page 11: ...manual for more information CAUTION New tips must be cleaned and sterilized high level disinfected before using them for the first time CAUTION A class B Vacuum autoclave according to EN 13060 is requ...

Page 12: ...iod of use IMPORTANT NOTICE All surfaces that the protection tip comes in contact with particularly the scanner tube and its optical window must always be cleaned and disinfected before attaching the...

Page 13: ...the correct rating and type 250 V T 3 15 A H must be used To replace the fuses first mount them in the drawer then insert the drawer back into the slot and push all the way in until the tabs click ba...

Page 14: ...3Shape TRIOS A S Holmens Kanal 7 1060 Copenhagen Denmark 10030970A April 2021 Copyright 3Shape TRIOS A S Let s change dentistry together More information at 3Shape com...