background image

9

ENGLISH

5.4 Electrical Safety

Ensure a grounded/earthed connection

WARNING

To avoid the risk of electric shock, the MOVE/MOVE+ cart 
must only be connected to a supply mains with grounding/
protective earth . The MOVE/MOVE+ cart is equipped with 
a three-pin (US: prongs) grounding/earth type plug . If you 
cannot insert the plug into your outlet, contact your elec-
trician to replace the plug or outlet, to ensure a grounded 
power connection . Do not try to defeat or ignore the 
purpose of the grounding-type plug .

Electrical shock

WARNING

There is a risk of electrical shock if you attempt to access 
the inside of any part of the system . Only authorized and 
qualified service personnel may access the inside of any 
part of the system .

Do not touch the electrical terminals 

WARNING

The smart cord, the scanner tip, and the scanner tip inter-
face surface on the scanner contain electrical terminals . 
For electrical safety, and to prevent corrosion and other 
damages, the electrical terminals must not be touched at 
any time, including touching electrical contacts and the 
patient at the same time .

Do not use a power strip

WARNING

Do not connect the MOVE/MOVE+ cart to an ordinary power 
strip or extension cord, because these may not be secure in 
relation to grounding . If the ground connection fails, these 
can cause the following hazards and potential injury:

- The total leakage current for all the connected equipment 
can exceed the limits specified in EN (IEC) 60601-1 .

- The impedance of the ground connection could exceed 
the limits specified in EN (IEC) 60601-1 .

Stress on cables

CAUTION

All externally connected cables (the power cable,  
network cable, and the smart cord) must never  
be subjected to pulling stress .

Spilled liquids

WARNING

Do not bring liquids, such as beverages, near the TRIOS IOS 
system . Do not spill liquids on the TRIOS IOS system .

Condensation

WARNING

Variations in temperature or humidity may cause water 
to condense inside the system, which can cause damage . 
Always let the system reach room temperature before you 
plug it into the power source .

If the system has been subject to major temperature or 
humidity changes, wait till the system has reached room 
temperature before plugging it into the power source .

If there is visible evidence of condensation, wait at least 8 
hours before plugging it into the power source, to ensure 
that no internal condensation occurs .

Disconnecting from mains

WARNING

There is no power ON/OFF switch on the system; therefore,  
the only reliable means to disconnect the system from 
mains is to unplug the power cord . Do not position the 
system so that it is difficult to unplug the power cord . 
Please allow enough time for the system to shut down 
properly before disconnecting the power cord .

System S3P (pod) only: You must also disconnect  
the PC from mains .

PC connection options and requirements 

System S3P (pod) only

WARNING

The PC is a standard IT equipment certified to IEC 60950-1 
and is located at least 1 .5 m away from the patient . If you 
choose this setup, you must not scan the patient and touch 
the PC at the same time! You must not touch any periph-
erals connected with a cable to the PC while scanning a 
patient . See the PC’s instructions for use for further details .

USB connection options and requirements

System S3A (MOVE/MOVE+ cart) only

WARNING

Only bus-powered USB devices without external power 
supply that support USB Full speed mode must be con-
nected to the USB port .

Summary of Contents for TRIOS S3A-10

Page 1: ...04 2021 3Shape TRIOS Safety and Setup Guide S3P 2 S3P 8 S3A 10 S3A 14 S3A 18 S3A 22 ...

Page 2: ...ormity EU 3ShapeTRIOSA SherebydeclaresthattheTRIOSobservestheprovisionsof Medical Device Regulation EU 2017 745 MDR Restriction of Hazardous Substances in Electrical and Electronic Equipment Directive 2011 65 EU RoHS Waste Electrical and Electronic Equipment Directive 2012 19 EU WEEE Registration Evaluation Authorization and Restriction of Chemicals Regulation 1907 2006 EC REACH Radio Equipment Di...

Page 3: ... Electronic Equipment Directive 2012 19 EU Radio Equipment Directive 2014 53 EU CAUTION EU only CE marking for connected products Further products which are connected to this unit must also bear the CE marking IMPORTANT NOTICE To comply with the CE marking of the integrated wireless module the wireless scanner must only be used indoors Electrical Safety This system has been tested to conform with ...

Page 4: ...lt in PC 3 scanner batteries and a scanner battery charger You can find the figures in the final section of this guide All system variants comprise Handheld TRIOS intraoral scanner S3 2 S1 and S2 Wired and wireless handheld dental intraoral scanner for scanning the surface of the patient s teeth TRIOS 4 SmartTip TST 01 S3 Re usable scanner tip attached to the handheld TRIOS Scanner The scanner tip...

Page 5: ...your service provider if there is visible damage or malfunctioning System S3A MOVE MOVE cart only Please follow the steps below for assembling TRIOS IOS system Figure 5 Step 1 Place the MOVE MOVE cart optimally in relation to the patient chair and your work position S24 Step 2 Attach the protection tip to the TRIOS scanner if not already in place S25 Step 3 Insert a battery fully into the rear of ...

Page 6: ...r is due Other messages include a reminder to mount the scanning tip Please make sure the scanning tip is mounted on the scanner Calibration A calibration of the handheld TRIOS scanner is required before using TRIOS for the first time or to compensate for thermal variation and aging effects due to wear During regular use the TRIOS software will prompt the user when calibration is due To perform th...

Page 7: ...ependent on local reg ulations and the configuration of the product 5 Safety and Warnings 5 1 Prerequisites CAUTION Read all instructions carefully including all warnings and caution notices You must comply with the warnings in this manual to prevent injury to persons and damage to equipment Proper functionality and safety can only be guaranteed if the safety precautions in this safety guide and o...

Page 8: ...d it should immediately be calibrated before further use If calibration fails please contact your technical service provider See this guide s instructions on calibrating the handheld scanner WARNING If the handheld scanner is dropped examine the TRIOS IOS system for any mechanical damage If a mechanical damage is observed or suspected treat this as an equip ment failure see this guide s instructio...

Page 9: ...ower cable network cable and the smart cord must never be subjected to pulling stress Spilled liquids WARNING Do not bring liquids such as beverages near the TRIOS IOS system Do not spill liquids on the TRIOS IOS system Condensation WARNING Variations in temperature or humidity may cause water to condense inside the system which can cause damage Always let the system reach room temperature before ...

Page 10: ... S3P pod only WARNING Before each patient make sure any PC connected to the system and its peripherals are disinfected It is recom mended to use medical grade peripherals e g keyboards and mice that can be disinfected easily 5 7 Safety when scanning with TRIOS CAUTION The handheld scanner is a highly precise optical instrument and must be handled carefully Do not drop shake or bump the handheld sc...

Page 11: ... manual for more information CAUTION New tips must be cleaned and sterilized high level disinfected before using them for the first time CAUTION A class B Vacuum autoclave according to EN 13060 is required for sterilization of the scanner tip CAUTION The scanner tip can be re sterilized up to 150 times and must be disposed afterwards as described in section Disposal CAUTION The scanner tip can be ...

Page 12: ...riod of use IMPORTANT NOTICE All surfaces that the protection tip comes in contact with particularly the scanner tube and its optical window must always be cleaned and disinfected before attaching the protection tip 6 4 Additional cleaning in case of poor scan quality During scanning the tip mirror may become unclean resulting in a loss of scan quality In this situation you can clean the mirror fo...

Page 13: ...h the correct rating and type 250 V T 3 15 A H must be used To replace the fuses first mount them in the drawer then insert the drawer back into the slot and push all the way in until the tabs click back into place and the drawer is no longer loose in the slot Do not attempt to insert the fuses directly into the slot on the MOVE MOVE cart The mains cable can now be reconnected and the TRIOS IOS sy...

Page 14: ...3Shape TRIOS A S Holmens Kanal 7 1060 Copenhagen Denmark 10030970A April 2021 Copyright 3Shape TRIOS A S Let s change dentistry together More information at 3Shape com ...

Reviews: