8. Technical Maintenance

CAUTION

All technical service and repair of the TRIOS IOS system 
must be performed by 3Shape or 3Shape-authorized 
personnel only .

The user doesn’t need to perform any regular maintenance other than  
calibration, cleaning, disinfection and sterilization, as described here . 
Preventive inspections and scheduled maintenance are not required .

Scanner batteries need to be disposed of as lithium-ion batteries .  
Lithium-ion batteries are recyclable and must not be treated as unsorted 
municipal waste, but must be collected separately!

Battery Replacements

Scanner Battery  
(S3P-2, S3A-14, S3A-22)

3Shape TB-02 Rechargeable Battery: 

 

3 .6 V, 2000mAh, 7 Wh

3Shape TB-03 Rechargeable Battery: 

 

3,6 V, 3450 mAh, 12 Wh

System S3A (MOVE/MOVE+ cart) only

Changing the fuses (Figure 2, S21):

Shut down the TRIOS MOVE/MOVE+ cart by following the procedure 
described in Setting up & Starting TRIOS for the First Time and disconnect 
the mains power cable from the TRIOS MOVE/MOVE+ cart .

CAUTION

The mains power cable must be disconnected  
before changing fuses .

Using a pair of tweezers or similar, push in the two tabs on each side of the 
fuse drawer simultaneously, as shown in the picture S21 . The drawer will 
begin to emerge from its slot .

It is then possible to remove the drawer and replace the two fuses .  
Two spare fuses are provided with the system on delivery .

WARNING

Only fuses with the correct rating and type  
(250 V, T 3 .15 A H) must be used .

To replace the fuses, first mount them in the drawer, then insert the drawer 
back into the slot and push all the way in until the tabs click back into place 
and the drawer is no longer loose in the slot . Do not attempt to insert the 
fuses directly into the slot on the MOVE/MOVE+ cart .

The mains cable can now be reconnected, and the TRIOS IOS system started up .

Changing the dongle for the MOVE/MOVE+ cart  
(software license protection):

The USB wireless dongle, the USB license dongle and the USB wifi dongle, and  
the Bluetooth dongle are located behind the plastic lid on top of the backpack  
of the MOVE/MOVE+ cart . You can remove the lid by opening the screws 
holding the lid, as indicated by the arrows in Figure 7 . 

WARNING

The scanner must not be operated while the ventilation 
grid on the back side of the MOVE/MOVE+ cart is removed .

9. Consumable Accessories

The consumable accessories to the TRIOS IOS System are the scanner tips, 
the scanner batteries, and the calibration kit .

When the scanner tips are disposed, new tips can be ordered by contacting 
your 3Shape reseller . Also, a new calibration  kit can be ordered from your 
3Shape reseller .

If the system is supplied with batteries, new batteries can be ordered  
by contacting your 3Shape reseller .

10. Vigilance

In case of any serious incident that has occurred in relation to the use of the 
device, please report the incident to 3Shape at email: support@3shape .com 
and you can also report it to the competent authority of the state in which 
the user and/or patient is established .

Summary of Contents for TRIOS S3A-10

Page 1: ...04 2021 3Shape TRIOS Safety and Setup Guide S3P 2 S3P 8 S3A 10 S3A 14 S3A 18 S3A 22 ...

Page 2: ...ormity EU 3ShapeTRIOSA SherebydeclaresthattheTRIOSobservestheprovisionsof Medical Device Regulation EU 2017 745 MDR Restriction of Hazardous Substances in Electrical and Electronic Equipment Directive 2011 65 EU RoHS Waste Electrical and Electronic Equipment Directive 2012 19 EU WEEE Registration Evaluation Authorization and Restriction of Chemicals Regulation 1907 2006 EC REACH Radio Equipment Di...

Page 3: ... Electronic Equipment Directive 2012 19 EU Radio Equipment Directive 2014 53 EU CAUTION EU only CE marking for connected products Further products which are connected to this unit must also bear the CE marking IMPORTANT NOTICE To comply with the CE marking of the integrated wireless module the wireless scanner must only be used indoors Electrical Safety This system has been tested to conform with ...

Page 4: ...lt in PC 3 scanner batteries and a scanner battery charger You can find the figures in the final section of this guide All system variants comprise Handheld TRIOS intraoral scanner S3 2 S1 and S2 Wired and wireless handheld dental intraoral scanner for scanning the surface of the patient s teeth TRIOS 4 SmartTip TST 01 S3 Re usable scanner tip attached to the handheld TRIOS Scanner The scanner tip...

Page 5: ...your service provider if there is visible damage or malfunctioning System S3A MOVE MOVE cart only Please follow the steps below for assembling TRIOS IOS system Figure 5 Step 1 Place the MOVE MOVE cart optimally in relation to the patient chair and your work position S24 Step 2 Attach the protection tip to the TRIOS scanner if not already in place S25 Step 3 Insert a battery fully into the rear of ...

Page 6: ...r is due Other messages include a reminder to mount the scanning tip Please make sure the scanning tip is mounted on the scanner Calibration A calibration of the handheld TRIOS scanner is required before using TRIOS for the first time or to compensate for thermal variation and aging effects due to wear During regular use the TRIOS software will prompt the user when calibration is due To perform th...

Page 7: ...ependent on local reg ulations and the configuration of the product 5 Safety and Warnings 5 1 Prerequisites CAUTION Read all instructions carefully including all warnings and caution notices You must comply with the warnings in this manual to prevent injury to persons and damage to equipment Proper functionality and safety can only be guaranteed if the safety precautions in this safety guide and o...

Page 8: ...d it should immediately be calibrated before further use If calibration fails please contact your technical service provider See this guide s instructions on calibrating the handheld scanner WARNING If the handheld scanner is dropped examine the TRIOS IOS system for any mechanical damage If a mechanical damage is observed or suspected treat this as an equip ment failure see this guide s instructio...

Page 9: ...ower cable network cable and the smart cord must never be subjected to pulling stress Spilled liquids WARNING Do not bring liquids such as beverages near the TRIOS IOS system Do not spill liquids on the TRIOS IOS system Condensation WARNING Variations in temperature or humidity may cause water to condense inside the system which can cause damage Always let the system reach room temperature before ...

Page 10: ... S3P pod only WARNING Before each patient make sure any PC connected to the system and its peripherals are disinfected It is recom mended to use medical grade peripherals e g keyboards and mice that can be disinfected easily 5 7 Safety when scanning with TRIOS CAUTION The handheld scanner is a highly precise optical instrument and must be handled carefully Do not drop shake or bump the handheld sc...

Page 11: ... manual for more information CAUTION New tips must be cleaned and sterilized high level disinfected before using them for the first time CAUTION A class B Vacuum autoclave according to EN 13060 is required for sterilization of the scanner tip CAUTION The scanner tip can be re sterilized up to 150 times and must be disposed afterwards as described in section Disposal CAUTION The scanner tip can be ...

Page 12: ...riod of use IMPORTANT NOTICE All surfaces that the protection tip comes in contact with particularly the scanner tube and its optical window must always be cleaned and disinfected before attaching the protection tip 6 4 Additional cleaning in case of poor scan quality During scanning the tip mirror may become unclean resulting in a loss of scan quality In this situation you can clean the mirror fo...

Page 13: ...h the correct rating and type 250 V T 3 15 A H must be used To replace the fuses first mount them in the drawer then insert the drawer back into the slot and push all the way in until the tabs click back into place and the drawer is no longer loose in the slot Do not attempt to insert the fuses directly into the slot on the MOVE MOVE cart The mains cable can now be reconnected and the TRIOS IOS sy...

Page 14: ...3Shape TRIOS A S Holmens Kanal 7 1060 Copenhagen Denmark 10030970A April 2021 Copyright 3Shape TRIOS A S Let s change dentistry together More information at 3Shape com ...

Reviews: