For transmitters rated at a maximum output power not listed above, the recommended 
separation distance 

d

 in metres (m) can be estimated using the equation applicable  

to the frequency of the trans-mitter, where 

P

 is the maximum output power rating  

of the transmitter in watts (W) according to the transmitter manufacturer.

Note 1:

 At 80 MHz and 800 MHz, the separation distance for the higher frequency  

range applies.

Note 2:

 These guidelines may not apply in all situations. Electromagnetic propagation  

is affected by absorption and reflection from structures, objects and people.

Models L1P-1 & L1P-1F

Connection

Type

Recommended use distances

Scanner – PC 

Wireless 

Distance between TRIOS scanner and wifi dongle/
access point should be between 0.5 and 5m for best 
performance.

5. System Safety

5.1 Prerequisites

CAUTION

Read all instructions carefully, including all warnings and caution notices. 
You must comply with the warnings in this manual to prevent injury  
to persons and damage to equipment. Proper functionality and safety can 
only be guaranteed if the safety precautions in this safety guide and on  
the TRIOS system are observed.

For L1P-1F: 

CAUTION

In fluorescence scanning mode:
•  Limit exposure of the oral mucosa.

Preventive inspection before use of the system

WARNING

Please examine the TRIOS system for any mechanical damage on:
•  All enclosures.
•  All cables.
Safety can only be guaranteed if NO DAMAGE to the TRIOS system is observed.

Modifications of the system

Warning

Do not modify the TRIOS system without authorization of the manufacturer. 

Approved software only

CAUTION

Install only approved software – to prevent interference with TRIOS system 
performance.   

Only connect specified items

Warning

Only connect items that have been specified as part of the TRIOS system  
or have been specified as being compatible with the TRIOS system. 

Training

WARNING

Before you attempt to use the TRIOS system with patients:
•  You must have read and understood all sections of this guide describing 

correct operation.

•  You must be thoroughly familiar with the safe operation of the TRIOS 

system as described in this documentation.

•  If the instructions for operating the TRIOS system as described in this 

guide are NOT observed and followed, the intended function and its 
safety for the user and patient may be impaired.

In case of equipment failure 

CAUTION

If at any time the TRIOS system malfunctions, or if you suspect in any way 
that the TRIOS system is not working correctly:
•  Remove the TRIOS scanner contact to patient immediately. 
•  Unplug the charger and make sure it cannot be used before it is checked.
•  Contact your 3Shape support.
•  Do not attempt to open any covers on the TRIOS system.

Wireless Connection

The wireless TRIOS scanner is intended to be used indoors, in a domestic or commercial 
electromagnetic environment. The wireless connection is intended to relay image stream  
data and control data between the TRIOS scanner and the TRIOS software. The communication 
between TRIOS Scanner and TRIOS software is handled by 3Shape communication services.

CAUTION

Other portable and mobile RF communication equipment could interfere 
with the wireless TRIOS scanner, even if the other equipment complies with 
CISPR emission requirements.

CAUTION

The wireless TRIOS scanner could interfere with other portable and mobile 
RF communication equipment.

IMPORTANT NOTICE

It is recommended to keep a line of sight between the wireless TRIOS 
scanner and the USB wireless dongle attached to the PC.

5.2 Potential Mechanical Damages

Inspect TRIOS scanner tip before using

WARNING

Damaged tip may cause injuries to the patient. 
Inspect TRIOS scanner tips before use to ensure 
that there is no damage to the TRIOS scanner  
tip surfaces. 

Dropped or damaged equipment

WARNING

Performance of the scanner tip is compromised 
when dropped. If the TRIOS scanner tip is dropped 
on the floor, the tip must be disposed of and not  
to be used for scanning. (see this guide’s instructions 
on 9. Disposal). 

WARNING

The performance of  the TRIOS scanner may be 
compromised when dropped. If the handheld 
scanner is dropped or bumped, it must be 
checked immediately before further use. 
•  Unplug the system and make sure it cannot be 

used before it is checked. 

•  Examine the TRIOS system for any mechanical 

damage or malfunction.

•  If at any time the TRIOS system malfunctions,  

or if you suspect in any way that the TRIOS 
system is not working correctly contact your 
reseller or 3Shape. 

•  Do not attempt to open any covers on the  

TRIOS system.

WARNING

The performance of the battery may be 
compromised when dropped, If a battery is 
dropped, It must be examined for any mechanical 
damage or malfunction. If a mechanical damage 
or malfunction is observed or suspected, the 
battery must be disposed of (see this guide’s 
instructions on 9. Disposal).

WARNING

The performance of the battery charger 
may be compromised when dropped, If the 
battery charger is dropped, it must be checked 
immediately before further use:
•  Unplug the TRIOS system and make sure it cannot 

be used before it is checked. 

•  Battery charger must be examined for any 

mechanical damages or malfunction. 

•  If mechanical damage or malfunction 

is observed, please contact support 
([email protected]).

WARNING

To avoid injuries from falling object, always place the TRIOS system on  
a horizontal surface.

5.3 Explosion Hazards 

Flammable liquids or gases

WARNING

The TRIOS system is not designed to be used in potentially explosive 
environments. Do not use the TRIOS system in close proximity to 
flammable liquids or gaseous flammable anaesthetic, or in oxygen-
enriched atmospheres.

Summary of Contents for L1P-1

Page 1: ...APR 2022 3Shape TRIOS 5 Safety and Setup Guide L1P 1 L1P 1F...

Page 2: ...system please carefully read and follow the instructions in this safety and setup guide and in the online helpsystem Pleasealwayskeepthissafetyandsetupguidenearthesystem Please follow all safety infor...

Page 3: ...safety requirements for medical electrical equipment by combining it with other equipment is responsible for conducting appropriate inspection and testing to ensure continued safe use of the equipment...

Page 4: ...Systembe performed byTRIOSauthorizedtechnicalservicepersonnel It is recommended to fully charge batteries for the TRIOS scanner before use CAUTION Let the system adjust to room temperature before use...

Page 5: ...midity 10 85 non condensing Atmospheric pressure 800 1100 hPa Transport Conditions Ambient temperature range 10 50 C 14 122 F Relative humidity 10 85 Non condensing Atmospheric pressure 620 1200 hPa 4...

Page 6: ...e b V1 3 Vrms V2 6 Vrms Note b Radiated RF IEC 61000 4 3 3 V m 80 MHz to 2 7 GHz E1 3 V m 80 MHz to 2 7 GHz Electromagnetic environment guidelines Portable and mobile RF communications equipment shoul...

Page 7: ...Connection The wireless TRIOS scanner is intended to be used indoors in a domestic or commercial electromagnetic environment The wireless connection is intended to relay image stream dataandcontrolda...

Page 8: ...ion in visual acuity and afterimages leading to irritation annoyance visual impairment and even accidents depending on the situation 5 6 Hygiene Forinstructionsonhowtomountthesingle usesleeveseetheTRI...

Page 9: ...nner during patient examination WARNING The TRIOS Sleeve must only be used once to avoid cross contamination 7 Cleaning Disinfection and Sterilization The following TRIOS system items shall be cleaned...

Page 10: ...ants Dispose of the CaviWipe Step 3 Remove gross contaminants from metal holder Take a CaviWipe Wipe the surface of the metal holder to remove any gross contaminants Dispose of the CaviWipe Step 4 Dis...

Page 11: ...TRIOS scanner window follow these steps for optimal scan quality Step 1 Take a damp non woven swab and gently rub it on the inside and outside of the TRIOS scanner tip window as well as on the window...

Page 12: ...10030138A April 2022 Copyright 3Shape TRIOS A S More information at 3Shape com Let s change dentistry together 3Shape TRIOS A S Holmens Kanal 7 1060 Copenhagen Denmark...

Reviews: