Dispose of the CaviWipe.

Step 4:

 Disinfect

Take a CaviWipe.

Wet the surface of the TRIOS scanner body and 
mirror housing with the CaviWipe for a minimum  
of 120 seconds to disinfect. 

Allow to dry.

Dispose of the CaviWipe.

Place TRIOS protection tip on TRIOS scanner.

CAUTION

The TRIOS scanner’s window is a delicate optical component. Its clean and 
undamaged surface is critical for the scan quality. The window should be 
handled with care. Make sure it is free from lint, stains, and other kinds of dirt.

CAUTION

Never submerge the TRIOS scanner into any cleaning or disinfection solution.

CAUTION

Before returning the TRIOS scanner to its holder, ensure both are dry. 

IMPORTANT NOTICE

The TRIOS scanner shell might get yellow-toned over a long period of use.

7.2 Pod: Cleaning and Disinfection

It is recommended to disassemble the pod in order to thoroughly clean it.  
Follow the instructions below (step 1-6)

Step 1:

 Disassemble pod

Wearing gloves, remove the rubber insert from  
the pod (holder) by pressing down on the middle 
using a thumb and gripping the outer edge with  
the index finger. 

Step 2:

 Remove gross contaminants from rubber insert

Take a CaviWipe.

Wipe the surface of the top rubber insert to remove 
any gross contaminants.

Dispose of the CaviWipe.

Step 3:

 Remove gross contaminants from metal holder

Take a CaviWipe.

Wipe the surface of the metal holder to remove any 
gross contaminants. 

Dispose of the CaviWipe.

Step 4:

 Disinfect rubber insert

Take another CaviWipe.

Wet the surface of the rubber insert for minimum 
120 seconds to disinfect. 

Dispose of the CaviWipe.

Step 5:

 Disinfect metal holder

Take a CaviWipe.

Wet the surface of the metal part for minimum  
120 seconds to disinfect.

Dispose of the CaviWipe.

Step 6:

 Assemble pod

Allow both parts to dry. Assemble the pod (holder).

7.3 TRIOS Scanner Tip: Cleaning and Sterilization

The TRIOS scanner tip is the part of the TRIOS scanner that is inserted into the patient’s 
mouth during scanning. The TRIOS scanner tip can be re-used up to 60 times. The TRIOS 
scanner tip should be cleaned and sterilized prior to its first use on patients and after each 
use on a patient to avoid cross-contamination.

Step 1:

 Clean TRIOS scanner tip

Wearing gloves, clean the TRIOS scanner tip 
manually using detergent diluted according  
to manufacturer’s instructions.

Apply the detergent directly on a soft bristled brush. 

Wash the TRIOS scanner tip under running lukewarm 
(30-40 °C) utility water.  

Inspect the window of the TRIOS scanner tip 
after cleaning. If the window appears with stains, 
smudged or with a milky haze, repeat this step.

Step 2:

 Rinse and dry TRIOS scanner tip

Rinse with lukewarm (30-40 °C) tap water.

Summary of Contents for L1P-1

Page 1: ...APR 2022 3Shape TRIOS 5 Safety and Setup Guide L1P 1 L1P 1F...

Page 2: ...system please carefully read and follow the instructions in this safety and setup guide and in the online helpsystem Pleasealwayskeepthissafetyandsetupguidenearthesystem Please follow all safety infor...

Page 3: ...safety requirements for medical electrical equipment by combining it with other equipment is responsible for conducting appropriate inspection and testing to ensure continued safe use of the equipment...

Page 4: ...Systembe performed byTRIOSauthorizedtechnicalservicepersonnel It is recommended to fully charge batteries for the TRIOS scanner before use CAUTION Let the system adjust to room temperature before use...

Page 5: ...midity 10 85 non condensing Atmospheric pressure 800 1100 hPa Transport Conditions Ambient temperature range 10 50 C 14 122 F Relative humidity 10 85 Non condensing Atmospheric pressure 620 1200 hPa 4...

Page 6: ...e b V1 3 Vrms V2 6 Vrms Note b Radiated RF IEC 61000 4 3 3 V m 80 MHz to 2 7 GHz E1 3 V m 80 MHz to 2 7 GHz Electromagnetic environment guidelines Portable and mobile RF communications equipment shoul...

Page 7: ...Connection The wireless TRIOS scanner is intended to be used indoors in a domestic or commercial electromagnetic environment The wireless connection is intended to relay image stream dataandcontrolda...

Page 8: ...ion in visual acuity and afterimages leading to irritation annoyance visual impairment and even accidents depending on the situation 5 6 Hygiene Forinstructionsonhowtomountthesingle usesleeveseetheTRI...

Page 9: ...nner during patient examination WARNING The TRIOS Sleeve must only be used once to avoid cross contamination 7 Cleaning Disinfection and Sterilization The following TRIOS system items shall be cleaned...

Page 10: ...ants Dispose of the CaviWipe Step 3 Remove gross contaminants from metal holder Take a CaviWipe Wipe the surface of the metal holder to remove any gross contaminants Dispose of the CaviWipe Step 4 Dis...

Page 11: ...TRIOS scanner window follow these steps for optimal scan quality Step 1 Take a damp non woven swab and gently rub it on the inside and outside of the TRIOS scanner tip window as well as on the window...

Page 12: ...10030138A April 2022 Copyright 3Shape TRIOS A S More information at 3Shape com Let s change dentistry together 3Shape TRIOS A S Holmens Kanal 7 1060 Copenhagen Denmark...

Reviews: