Level 6500
This code stands for disconnection of the tubes. It could also occur after you cut the tubes so,
first of all, try and recalibrate the sensor.
Fig. 89
Level 6500
Fig. 90
CNT below 10 000
If doesn’t help, check the CNT. If it is below 10 000, then most likely the tubes are
disconnected from the sensor’s head indeed.
Take the picture of the sensor’s head (the serial number must be clearly visible), its tubes (it
must be clearly visible if there is any misalignment of the tubes) and the screenshot of the main
screen of the sensor and of the Settings menu and send all those to us.