64220, Rev. 1/April 2019
3M™ GMI™ PS200-LMP AUTO BUMP/CALIBRATION STATION
RE-CONFIGURING THE ABC
7-2
4. Select
Bump Test
or
Calibration
as applicable (see
).
Figure 7-3: Select Bump Test or Calibrate
5. Enter the
Gas 1
and
Gas 2
concentrations as detailed on the gas cylinder labels. The example in
illustrates Gas 1 concentrations of 2.4% Methane and 95 ppm
Carbon Monoxide.
Figure 7-4: Select Gas Concentrations
6. Click
Generate Settings File
. This creates a
new ABC configuration file on the USB memory stick.