2

Important Safety Information

3M™ Clean‑Trace Luminometer LX25

Please read, understand, and follow all safety information

contained in these instructions prior to the use of this

3M™ Clean‑Trace™ Luminometer. Retain these instructions

for future reference.

Product Description and Intended Use:

The 3M™ Clean-Trace™ Luminometer LX25 is a battery operated

Luminometer utilized primarily for objective and sensitive

determination of the end result of 3M™ Clean-Trace™ Adenosine

Tri-Phosphate (ATP) tests . The Clean-Trace Luminometer LX25

power supply is provided to charge the battery .
3M has not designed nor documented the Clean-Trace

Luminometer LX25 or Clean-Trace Luminometer LX25 power

supply for use with other manufacturers’ products, use with

other manufacturers’ designs, or use in any other application

that has not been evaluated or documented by 3M and may lead

to an unsafe condition . The Clean-Trace Luminometer LX25

is not intended to be used in potentially flammable or

explosive environments . Surfaces or solutions should not be

considered sterile based on results from the 3M Clean-Trace

Luminometer LX25 . The 3M Clean-Trace Luminometer LX25 is

not intended to be used in the diagnosis of conditions in humans

or animals . The user is responsible to know and follow applicable

workplace regulations . Clean-Trace ATP tests, equipment, and

software must be utilized by technicians that have been properly

trained on the 3M™ Clean-Trace™ ATP Monitoring System

according to all applicable instructions for use .
If this product is used in a manner not specified, the results

provided by the product may be impaired .

Explanation of Signal Word Consequences

WARNING:

 Indicates a hazardous situation which, if

not avoided, could result in serious injury or death

CAUTION:

 Indicates a hazardous situation which, if

not avoided, could result in minor or moderate injury

and/or property damage .

NOTICE:

Indicates a situation which, if not avoided, could

result in property damage .

Explanation of Safety Related Symbols

Warning: Hazardous Voltage

Attention: Read Accompanying Documentation

CAUTION: Recycle 3M™ Clean-Trace

Luminometer LX25 to avoid environmental

contamination . This product contains recyclable

parts . For information on recycling, in the US

please contact 3M Health Care Service Center

at 1-800-292-6298 for additional information .

Outside of the US contact your local 3M office .

Li

WEEE mark - This product contains electrical

and electronic components and must not be

disposed of using standard refuse collection .

Please consult local directives for disposal of

electrical and electronic equipment .

Class II: Class two insulation referred to as

double insulation .

Applies to 3M Clean-Trace

Luminometer LX25 Power Supply: Alternating

current AC source .

Applies to 3M Clean-Trace

Luminometer LX25 Power Supply: Direct

current DC source .

UL NRTL mark (US and Canada electrical safety)

CE mark (conforms to applicable requirements

for European Economic Area) .

RCM mark (Australia electrical safety and EMC)

BC mark (California regulation for battery

charging systems)

REF

The 3M Clean-Trace Luminometer LX25 model

number, which is LX25 .

The 3M Clean-Trace

Luminometer LX25 serial number

Efficiency: Complies to section

301 of Energy Independence

and Security Act (EISA)

complies with Energy Star tier .
2 (North America), ECP

tier 2 (China), MEPS tier

2 (Australia), Code of

Conduct (Europe)

Description:

Base File Name:

Software:
SGS Contact:
Email:
Printer:
Printer Location:
Supplier:

3M™ Clean‑Trace™

Luminometer LX25 User Manual

Requester: Carol Engdahl

Reference: LX25

Die # / Doc. Size: 8.5" x 11"

Item Spec#: 34-8720-2654-6

Supersedes#:

Cat/Product#:

Structure#:

Insert

PA:  LX25_34872026546.indd
GA:

InDesign CC 2014

PROCESS

BLACK

PROCESS

CYAN

PROCESS

MAGENTA

PROCESS

YELLOW

4567348

4

04.17.17 New Build

kmh

04.24.17 Alts per Marked up PDF and job ticket

kmh

05.26.17 Alts per Marked up PDF

kmh

06.12.17 Update images in the Chart.

Jer

Summary of Contents for Clean-Trace Luminometer LX25

Page 1: ...ETER LX25 7 HOW TO PERFORM A QUICK TEST 7 COMMUNICATING WITH THE 3M QUALITY CONTROL DATA MANGER QCDM 7 REGISTER A CLEAN TRACE LUMINOMETER LX25 TO AN ORGANIZATION ON QCDM 7 ROUTINE SYNCHRONIZATION 8 DE REGISTRATION 8 WI FI SET UP 9 VERIFYING DATA UPLOAD TO QCDM 9 APPLICATION WORKFLOW 9 STARTING THE AUDIT 9 HOME SCREEN 10 1 FLEXIBLE ENDOSCOPE WORKFLOW 10 2 ENVIRONMENTAL SURFACES WORKFLOW 11 3 SURGIC...

Page 2: ...ided by the product may be impaired Explanation of Signal Word Consequences WARNING Indicates a hazardous situation which if not avoided could result in serious injury or death CAUTION Indicates a hazardous situation which if not avoided could result in minor or moderate injury and or property damage NOTICE Indicates a situation which if not avoided could result in property damage Explanation of S...

Page 3: ...components with 3M designated replacement parts only To reduce the risks associated with illness or infection Do not use the Clean Trace Luminometer LX25 and 3M Clean Trace ATP Tests in the diagnosis of conditions in humans or animals Do not use the Clean Trace Luminometer LX25 and 3M Clean Trace ATP Tests as an indication of disinfection or sterilization of surfaces or devices CAUTION To reduce t...

Page 4: ...lly inspect the Clean Trace Luminometer LX25 for damage including cracks or swelling Do not store or use the Clean Trace Luminometer LX25 or power supply in a location exposed to steam high humidity dust or strong vibration Reference the Environmental Operating Conditions table in the Important Safety Information for Clean Trace Luminometer LX25 in this User Manual Use only 3M Clean Trace ATP Test...

Page 5: ...ntation THIS WARRANTY IS VOID IF A THE HARDWARE HAS BEEN REPAIRED BY PERSONS NOT AUTHORIZED BY 3M OR B THE HARDWARE HAS BEEN ALTERED MODIFIED OR MISUSED OR C THE HARDWARE IS USED WITH PRODUCTS SUPPLIES COMPONENTS OR SOFTWARE NOT SUPPLIED BY 3M FOR USE WITH THE HARDWARE OR D THE HARDWARE OR A COMPONENT IS USED FOR OTHER USES FOR EXAMPLE USE WITH OTHER CIRCUIT BOARDS OR SOFTWARE OR E THE HARDWARE HA...

Page 6: ...ing less than 10 charge capacity The battery does not need to be fully discharged before recharging When storing a Clean Trace Luminometer LX25 for more than a month the battery level should be at 50 or less charge Do not store a Clean Trace Luminometer LX25 with a fully discharged battery because the battery may lose the ability to be recharged Battery replacement is recommended after two years o...

Page 7: ...n Trace ATP Test UXC or H2O Instructions for Use to collect the sample activate the ATP Test and insert into the Clean Trace Luminometer LX25 Tap Start to take a measurement Remove the ATP Test UXC or H2O after the measurement is complete All results should be recorded o Only the last 20 Quick Tests are saved on the Clean Trace Luminometer LX25 o The results of a quick test are not saved to the Qu...

Page 8: ...sing the USB cable 5 After a short period a red spinning circle will appear on the Clean Trace Luminometer LX25 indicating the Clean Trace Luminometer LX25 is connected to the Synchronization Manager 6 On the workstation log into Synchronization Manager Software 7 Select the Start Synchronization button on the workstation Synchronization Manager Software 8 Start the Synchronization on the workstat...

Page 9: ...d An unsuccessful upload will result in a triangle If using the Synchronization Manager Software to upload results refer to section on Routine Synchronization of Results in this document All results will be marked with a triangle until LX25 results are synchronized and data is uploaded to the QCDM software Application Workflow Before access to the workflow screens is available an organization hier...

Page 10: ...Workflow 3 Surgical Instrument Workflow 1 Flexible Endoscope Workflow From the Home screen tap Start to begin new test or select a recent failed audit for retest Select device type Select Device This is the serial number or other unique identifier Select Reprocessing Step Review selections o Select technician o Can change device reprocessing step and test plan o Tap Perform Audit Select test point...

Page 11: ... Select Room Review Selections o Can change Room number test plan o Select Housekeeper and Isolation room option o Tap Perform Audit Choose test point Insert ATP Test UXC or H2O Press Begin Measurement and wait until blue Reading Result screen disappears then press Return Repeat process until all test points have been tested Press Done ...

Page 12: ...lt screen disappears then press Return Repeat process until all test points have been tested Press Done Performing a Retest To retest a failed test point From the home screen select the audit that requires retesting marked with a red X Select the test point that requires retesting marked with a red X Follow the prompts and select any additional test points if required Complete the retest Select Do...

Page 13: ... Clean Trace Luminometer LX25 System Diagnostics tap to initiate device diagnostic test Settings menu Settings Menu Functions detail o Wi Fi Settings turn Wi Fi on off set up network o Server settings select or change server connection o Screen brightness increase or decrease screen brightness o Screen timeout choose length of time before screen goes to sleep o Date set current date o Time set cur...

Page 14: ...an Trace ATP Test UXC or H2O has been left in the chamber for more than 30 seconds following the completion of a measurement If this situation is detected the instrument will display a Swab Left In Chamber error message on the screen and provide an audible beep to remind the user to remove the Test following the measurement No Display Battery fully discharged Recharge the battery Clean Trace Lumin...

Page 15: ...Clean Trace Luminometer LX25 Make sure the towel used is just damp not dripping 3 When cleaning the Clean Trace Luminometer LX25 avoid the USB cable connection on the side of the Clean Trace Luminometer LX25 4 Rinse disinfectant from Clean Trace Luminometer LX25 by repeating the procedure 2 3 using another disposable towel slightly dampened with distilled or deionized water 5 Allow Clean Trace Lum...

Page 16: ... St Paul MN 55144 USA www 3M com Medical 3M and Clean Trace are trademarks of 3M Used under license in Canada 2017 3M All rights reserved Issue Date 2017 04 34 8720 2654 6 3M Deutschland GmbH Health Care Business Carl Schurz Str 1 41453 Neuss Germany ...

Reviews: