Inserting a Clean‑Trace ATP Test UXC or H2O into the

Clean‑Trace Luminometer LX25

• Press the rocker cap to open the test chamber (1) .
• Insert an activated Clean-Trace ATP Test UXC or H2O (2) .
• The Test handle should be visible at the top of the

chamber (3) .

• After the reading is complete, press the rocker cap to open

the test chamber and remove the ATP Test UXC or H2O .

• If an ATP Test UXC or H2O remains in the Clean-Trace

Luminometer LX25 chamber for more than 30 seconds,

the Swab Sensor notifies with an audible tone .

Keep the Clean‑Trace Luminometer LX25 upright

whenever a test device is in the chamber, to ensure

accurate results and avoid spillage of reagents, which

could damage the sensitive components of the instrument .

• If the Clean-Trace Luminometer LX25 is held at greater

than 68 degrees from vertical when a test is in the

chamber, a message notifies the user that the Clean-Trace

Luminometer LX25 is in the incorrect position for testing

and to use a more vertical position when testing .

o

Tap “OK” then hold the Clean-Trace

Luminometer LX25 in a vertical position and repeat

the measurement .

How to Perform a Quick Test

A quick test is performed under circumstances where results are

not uploaded to the Quality Control Data Manager (e .g . training,

demonstrations, no access to Quality Control Data Manager) .

To perform a Quick Test

• Power on the Clean-Trace

Luminometer LX25 .

• After successful completion of

System Diagnostics this landing

screen with appear:

• Log in is not required to perform a

Quick Test .

• Tap “Quick Test”
• Follow the Clean-Trace ATP Test

UXC or H2O Instructions for Use to

collect the sample, activate the ATP

Test and insert into the Clean-Trace

Luminometer LX25 .

• Tap “Start” to take a measurement .
• Remove the ATP Test UXC or H2O after the measurement

is complete .

• All results should be recorded .

o

Only the last 20 Quick Tests are saved on the

Clean-Trace Luminometer LX25 .

o

The results of a quick test are not saved to the Quality

Control Data Manager .

Communicating with the

3M Quality Control Data Manger (QCDM)

Before the Clean-Trace Luminometer LX25 can be used with

the Quality Control Data Manager (QCDM) it must first be

synchronized to the facility’s QCDM organization .

To register the Clean-Trace Luminometer LX25, the facility

account must be linked to the facility’s organization and

the 3M™ Synchronization Manager Software installed on a

workstation (desktop/laptop) . Registration cannot be performed

using the Clean-Trace Luminometer LX25 Wi-Fi function .

Register a Clean‑Trace Luminometer LX25

to an Organization on QCDM

1 . Find the Synchronization Manager icon in system tray,

right click on the icon and choose “Show window” .

2 . Power up the Clean-Trace Luminometer LX25 and launch

the app if not already launched .

a . Follow the steps in the set-up

menus for language, date, time

and time zone .

b . Skip Wi-Fi set up .
c . Wait for diagnostics

to complete .

d . When the landing screen

appears -

STOP – remain on

landing screen – DO NOT

LOG IN.

3 . Using the USB cable provided, connect the Clean-Trace

Luminometer LX25 to the workstation running the

Synchronization Manager software .

4 . After a short period, a red, spinning

circle appears on the Clean-Trace

Luminometer LX25 screen,

indicating your Clean-Trace

Luminometer LX25 is connected

to the Synchronization

Manager Software .

Summary of Contents for Clean-Trace Luminometer LX25

Page 1: ...ETER LX25 7 HOW TO PERFORM A QUICK TEST 7 COMMUNICATING WITH THE 3M QUALITY CONTROL DATA MANGER QCDM 7 REGISTER A CLEAN TRACE LUMINOMETER LX25 TO AN ORGANIZATION ON QCDM 7 ROUTINE SYNCHRONIZATION 8 DE REGISTRATION 8 WI FI SET UP 9 VERIFYING DATA UPLOAD TO QCDM 9 APPLICATION WORKFLOW 9 STARTING THE AUDIT 9 HOME SCREEN 10 1 FLEXIBLE ENDOSCOPE WORKFLOW 10 2 ENVIRONMENTAL SURFACES WORKFLOW 11 3 SURGIC...

Page 2: ...ided by the product may be impaired Explanation of Signal Word Consequences WARNING Indicates a hazardous situation which if not avoided could result in serious injury or death CAUTION Indicates a hazardous situation which if not avoided could result in minor or moderate injury and or property damage NOTICE Indicates a situation which if not avoided could result in property damage Explanation of S...

Page 3: ...components with 3M designated replacement parts only To reduce the risks associated with illness or infection Do not use the Clean Trace Luminometer LX25 and 3M Clean Trace ATP Tests in the diagnosis of conditions in humans or animals Do not use the Clean Trace Luminometer LX25 and 3M Clean Trace ATP Tests as an indication of disinfection or sterilization of surfaces or devices CAUTION To reduce t...

Page 4: ...lly inspect the Clean Trace Luminometer LX25 for damage including cracks or swelling Do not store or use the Clean Trace Luminometer LX25 or power supply in a location exposed to steam high humidity dust or strong vibration Reference the Environmental Operating Conditions table in the Important Safety Information for Clean Trace Luminometer LX25 in this User Manual Use only 3M Clean Trace ATP Test...

Page 5: ...ntation THIS WARRANTY IS VOID IF A THE HARDWARE HAS BEEN REPAIRED BY PERSONS NOT AUTHORIZED BY 3M OR B THE HARDWARE HAS BEEN ALTERED MODIFIED OR MISUSED OR C THE HARDWARE IS USED WITH PRODUCTS SUPPLIES COMPONENTS OR SOFTWARE NOT SUPPLIED BY 3M FOR USE WITH THE HARDWARE OR D THE HARDWARE OR A COMPONENT IS USED FOR OTHER USES FOR EXAMPLE USE WITH OTHER CIRCUIT BOARDS OR SOFTWARE OR E THE HARDWARE HA...

Page 6: ...ing less than 10 charge capacity The battery does not need to be fully discharged before recharging When storing a Clean Trace Luminometer LX25 for more than a month the battery level should be at 50 or less charge Do not store a Clean Trace Luminometer LX25 with a fully discharged battery because the battery may lose the ability to be recharged Battery replacement is recommended after two years o...

Page 7: ...n Trace ATP Test UXC or H2O Instructions for Use to collect the sample activate the ATP Test and insert into the Clean Trace Luminometer LX25 Tap Start to take a measurement Remove the ATP Test UXC or H2O after the measurement is complete All results should be recorded o Only the last 20 Quick Tests are saved on the Clean Trace Luminometer LX25 o The results of a quick test are not saved to the Qu...

Page 8: ...sing the USB cable 5 After a short period a red spinning circle will appear on the Clean Trace Luminometer LX25 indicating the Clean Trace Luminometer LX25 is connected to the Synchronization Manager 6 On the workstation log into Synchronization Manager Software 7 Select the Start Synchronization button on the workstation Synchronization Manager Software 8 Start the Synchronization on the workstat...

Page 9: ...d An unsuccessful upload will result in a triangle If using the Synchronization Manager Software to upload results refer to section on Routine Synchronization of Results in this document All results will be marked with a triangle until LX25 results are synchronized and data is uploaded to the QCDM software Application Workflow Before access to the workflow screens is available an organization hier...

Page 10: ...Workflow 3 Surgical Instrument Workflow 1 Flexible Endoscope Workflow From the Home screen tap Start to begin new test or select a recent failed audit for retest Select device type Select Device This is the serial number or other unique identifier Select Reprocessing Step Review selections o Select technician o Can change device reprocessing step and test plan o Tap Perform Audit Select test point...

Page 11: ... Select Room Review Selections o Can change Room number test plan o Select Housekeeper and Isolation room option o Tap Perform Audit Choose test point Insert ATP Test UXC or H2O Press Begin Measurement and wait until blue Reading Result screen disappears then press Return Repeat process until all test points have been tested Press Done ...

Page 12: ...lt screen disappears then press Return Repeat process until all test points have been tested Press Done Performing a Retest To retest a failed test point From the home screen select the audit that requires retesting marked with a red X Select the test point that requires retesting marked with a red X Follow the prompts and select any additional test points if required Complete the retest Select Do...

Page 13: ... Clean Trace Luminometer LX25 System Diagnostics tap to initiate device diagnostic test Settings menu Settings Menu Functions detail o Wi Fi Settings turn Wi Fi on off set up network o Server settings select or change server connection o Screen brightness increase or decrease screen brightness o Screen timeout choose length of time before screen goes to sleep o Date set current date o Time set cur...

Page 14: ...an Trace ATP Test UXC or H2O has been left in the chamber for more than 30 seconds following the completion of a measurement If this situation is detected the instrument will display a Swab Left In Chamber error message on the screen and provide an audible beep to remind the user to remove the Test following the measurement No Display Battery fully discharged Recharge the battery Clean Trace Lumin...

Page 15: ...Clean Trace Luminometer LX25 Make sure the towel used is just damp not dripping 3 When cleaning the Clean Trace Luminometer LX25 avoid the USB cable connection on the side of the Clean Trace Luminometer LX25 4 Rinse disinfectant from Clean Trace Luminometer LX25 by repeating the procedure 2 3 using another disposable towel slightly dampened with distilled or deionized water 5 Allow Clean Trace Lum...

Page 16: ... St Paul MN 55144 USA www 3M com Medical 3M and Clean Trace are trademarks of 3M Used under license in Canada 2017 3M All rights reserved Issue Date 2017 04 34 8720 2654 6 3M Deutschland GmbH Health Care Business Carl Schurz Str 1 41453 Neuss Germany ...