2.  To span the monitor: 
 

a.  

Remove the protective plastic cap from the SPAN GAS cylinder, noting    

the concentration of carbon monoxide printed on the cylinder. 

 
b.

  Connect the plastic tubing provided in the kit to the plastic fitting, and 

connect the opposite end of the tubing securely to the barbed fitting on 
the regulator. 

 
c.

  Connect the monitor's sample tube to the proper end of the  

tubing/regulator assembly, described in the previous step. 

 
d.

  Locate the SPAN adjustment potentiometer and verify the monitor has 

had adequate warm-up time (see CO Monitor's 

User Instructions

). 

 

e.

  Screw the regulator valve onto the SPAN GAS cylinder outlet, making 

sure not to "cross thread" the regulator or cylinder outlet port, and 
tighten the regulator firmly. Open the knob on the regulator 
counterclockwise to start gas flow. 

 
f.

   Observe the display on the monitor. The reading will increase, and 

then stabilize. As soon as the reading is stable (approximately one 
minute), compare the value indicated on the display, with the value 
noted on the SPAN GAS cylinder. If the display does NOT read the 
same value as noted on the SPAN GAS cylinder, adjust the SPAN 
potentiometer until the reading indicated on the display matches the 
concentration of carbon monoxide printed on the SPAN GAS 
calibration cylinder. 

 
g.

  When these steps have been completed, close the knob on the 

regulator clockwise to stop the gas flow. Unscrew regulator valve from 
the SPAN GAS cylinder and replace the protective plastic cap. 

 
h. 

Disconnect the associated calibration test tubing from the monitor and 
reconnect the monitor's sample tube to the sampling supply on the 
filtration system. 

 

i. 

When using the small kit (part number 529-04-49 or automotive part 
number 72012), be aware that the regulator valve allows gas to flow 
from the cylinder as soon as it is tightened down.  To conserve the gas 
in the cylinder, do not attach the regulator valve until you are ready to 
begin spanning the monitor and remove the regulator valve as soon as 
spanning is complete.

 

 