A HOLD notice will also be
displayed to alert the user that the
instrument is currently in HOLD
status.  To unfreeze the display and
return to floating measurement,
momentarily press the HOLD switch
once again.   Please note that during
HOLD condition the distance-
indicating LED’s are turned off.

3.5 Zero adjustment: the 3M™ 718 Static

Sensor has a zero adjustment
function, which sets a zero reference
point for all subsequent
measurements.  This zero reference
can be set by pointing the instrument
at a known zero-voltage surface, and
holding down the RANGE/ZERO
button for longer than 3 seconds.
After 3 seconds, the display will flash
and adjust to zero.  Repeat this step
for both the 2 kV and the 20 kV
ranges.  The zero adjustment should
be performed every time the unit is
turned on.

3.6 Measurement Accuracy:

3.6.1 Distance Indicator: the 718 Static

Sensor is factory calibrated to give
accurate measurements when it is
placed one inch (2.54 cm) away
from the object to be measured.  To
assist the user in gauging this
distance, two light-emitting diodes
(LED’s) are present on the front face
of the instrument.  These LED’s
emit two red, bullseye targets on the
surface of the object being measured.
As the instrument gets closer to the
one inch measurement distance, the
bullseyes begin to converge.  When
they converge and become one, the
instrument is approximately one
inch away, and the measurement can
be made.  
For more accurate measurements, 
it is recommended that the user
manually measure the distance
between the front housing of the
instrument and the object 
being measured.

3.6.2 Accuracy and Size of Object to be

Measured: the minimum surface
area on an electrostatically charged
object which can be accurately
measured is a 5 square inch 
(32.3 cm

2

) area.

3.6.3 Measurements from Greater than

One Inch (2.54 cm) Away: in the
event that a one inch separation
between object-to-be measured and
the 718 cannot be achieved, it is
possible to get approximate
readings.  This can be accomplished
by holding the instrument at
multiples of one inch (two inches,
three inches, etc.) away from the
object.  The measurement reading is
then adjusted according to the
following equation:

Measurement distance (in inches) x 
displayed voltage (in kilovolts) = 
approximate voltage in kilovolts

Summary of Contents for 718

Page 1: ...fying the operation of air ionizers by measuring the neutralization time for a static charge placed on a stationary metal plate Any deviation from this intended use could impair the instrument s effec...

Page 2: ......

Page 3: ...lied by a 9V alkaline battery 2 2 To install the battery on either unit 2 2 1 Remove the battery cover located in the lower back of the unit To do this press down on the cover and slide it downward 2...

Page 4: ...ront of the sensor are lit To shut off the 718 Static Sensor momentarily press down on the membrane switch labeled POWER Determining that the power is off can be verified by seeing that the liquid cry...

Page 5: ...ight emitting diodes LED s are present on the front face of the instrument These LED s emit two red bullseye targets on the surface of the object being measured As the instrument gets closer to the on...

Page 6: ...disabled The Automatic Shut Off feature will reset itself the next time the instrument is turned on 4 0 Operation and Use of the 718A Air Ionizer Test Kit It is recommended that the user be familiar...

Page 7: ...er s operating manual or consult with the ionizer manufacturer to determine what this offset balance should be 5 0 Service Calibration 5 1 Service and Repair in the event that you believe the 718 Stat...

Page 8: ...he stand to ground 5 5 3 Turn on the 718 Static Sensor and set it to the 2 kV range 5 5 4 Ground the plate Zero the display 5 5 5 Remove the ground from the plate and connect it to the high voltage po...

Page 9: ...ge Automatic Shut off 20 minutes after last switch activity Operating Time for Battery Greater than 50 hours with new battery 21 C Continuous usage Distance Indicator LED targets Aligned targets indic...

Page 10: ...ual 1 ea Certificate of Conformance 8 0 Contact Information Customer and Technical Service Within the U S Customer service and technical support can be obtained by calling the 3M Electronic Handling P...

Page 11: ......

Page 12: ...less expressly agreed upon in writing by an authorized officer of 3M Warranty Limited Remedy Limited Liability This product will be free from defects in material and manufacture for a period of one ye...

Reviews: