4.3 Testing ionizer discharge time: after

charging the plate, hold the 718
approximately one foot (30.5 cm)
away from the ionizer.  Monitor the
display to see how quickly the 1.1 kV
charge is dissipated to 0.1 kV.  The
speed at which this occurs (the
discharge time) indicates how well
the ionizer is operating.  Please refer
to the specific ionizer’s operating
manual or consult with the ionizer
manufacturer to determine what this
discharge time should be.  Repeat this
procedure for both a positively and a
negatively charged plate.

4.4 Testing ionizer offset balance: zero

the charge plate by touching it with a
grounded object.  This can either be
the finger of a grounded person or
some other item which is connected
to electrical ground.  In either case,
zeroing the charge plate should make
the display on the 718 read zero.
Hold the 718 approximately one foot
(30.5 cm) in front of the ionizer.

Monitor the display.  The value
displayed is the offset balance of the
ionizer, which is the difference
between the number of positive and
negative ions being emitted. Please
refer to the specific ionizer’s
operating manual or consult with the
ionizer manufacturer to determine
what this offset balance should be.

5.0 Service/Calibration

5.1 Service and Repair: in the event that

you believe the 718 Static Sensor or
the  3M™ 718A Air Ionizer Test Kit
is in need of repair, please contact
your local 3M representative for
troubleshooting help, and, as needed,
repair information.  There are no
user-serviceable parts on either
product.

5.2 Calibration: the 718 and 718A

products are supplied by the factory
pre-calibrated.  3M does not specify a
minimum calibration cycle for the
718 or 718A products.  The user,
usually according to internal Quality
procedures, determines calibration
cycles.  3M does offer calibration
services on these instruments.  Please
contact your local 3M representative
for information on this service.  In the
event that the user wishes to perform
a self-calibration, the following steps
should be followed for the 718 Static
Sensor (user-calibration not possible
on the 718A Air Ionizer Test kit).

5.3 Equipment Needed:

Test fixture
High-Voltage Power Supply,

capable of supplying voltages up 
to 10,000 V

Voltmeter,

with > 50 k

Ω

input

impedance, capable of measuring
voltages down to the µV range.

Cable

with a 3/32 inch (2.5mm)

mono plug and secondary connector
to interface with voltmeter

Summary of Contents for 718

Page 1: ...fying the operation of air ionizers by measuring the neutralization time for a static charge placed on a stationary metal plate Any deviation from this intended use could impair the instrument s effec...

Page 2: ......

Page 3: ...lied by a 9V alkaline battery 2 2 To install the battery on either unit 2 2 1 Remove the battery cover located in the lower back of the unit To do this press down on the cover and slide it downward 2...

Page 4: ...ront of the sensor are lit To shut off the 718 Static Sensor momentarily press down on the membrane switch labeled POWER Determining that the power is off can be verified by seeing that the liquid cry...

Page 5: ...ight emitting diodes LED s are present on the front face of the instrument These LED s emit two red bullseye targets on the surface of the object being measured As the instrument gets closer to the on...

Page 6: ...disabled The Automatic Shut Off feature will reset itself the next time the instrument is turned on 4 0 Operation and Use of the 718A Air Ionizer Test Kit It is recommended that the user be familiar...

Page 7: ...er s operating manual or consult with the ionizer manufacturer to determine what this offset balance should be 5 0 Service Calibration 5 1 Service and Repair in the event that you believe the 718 Stat...

Page 8: ...he stand to ground 5 5 3 Turn on the 718 Static Sensor and set it to the 2 kV range 5 5 4 Ground the plate Zero the display 5 5 5 Remove the ground from the plate and connect it to the high voltage po...

Page 9: ...ge Automatic Shut off 20 minutes after last switch activity Operating Time for Battery Greater than 50 hours with new battery 21 C Continuous usage Distance Indicator LED targets Aligned targets indic...

Page 10: ...ual 1 ea Certificate of Conformance 8 0 Contact Information Customer and Technical Service Within the U S Customer service and technical support can be obtained by calling the 3M Electronic Handling P...

Page 11: ......

Page 12: ...less expressly agreed upon in writing by an authorized officer of 3M Warranty Limited Remedy Limited Liability This product will be free from defects in material and manufacture for a period of one ye...

Reviews: