![3DHISTECH P1000 Service Manual Download Page 84](http://html1.mh-extra.com/html/3dhistech/p1000/p1000_service-manual_442157084.webp)
P1000 SERVICE SOFTWARE MANUAL
14. SLIDE LOADING UNITS
SERVICE GUIDE
83
7.
Click the Setting sensors button. In this menu, you can fine tune the Slide edge sensor and the Slide out sensor. Also, you can define the position
of the Safety container here.
7.1.
Fine tuning the Slide edge sensor
If applicable, fill up four double magazines in a way that a slide is followed by an empty slot and a slide again. Fill up eight simple magazines
this way, too. Insert them to the magazine racks as illustrated.