In the right bar, click the
Stage control
panel. Move the Stage to a position so that the Stage and the granite slab markings are aligned as
illustrated above. Click the
Stage slide X
button again. Move the Arm to the Z-axis minimum to avoid possible collisions. Move the Arm on
into a position so that the longitudinal mark in the center of the slide is aligned with the mark on the Stage and the slab. Move the slide
closer to the Stage for better visibility. Click
on the side panel to set this X-position. Remove the slide from the Gripper.