Pannoramic MIDI 1.17 User’s Guide
5 Troubleshooting and Maintenance
•
When the HXP 120 illumination unit is switched on:
◦
NEVER
look directly into the fiber-optic cable.
◦
NEVER
remove the fiber-optic cable from the device.
◦
NEVER
connect the fiber-optic cable to the device.
◦
NEVER
remove the housing from the device.
5.1.1 Cleaning the device
•
Before cleaning, switch off the device and pull the plug out.
•
Do not let liquid penetrate into the system.
•
To clean liquid from the device, use a lint-free cotton cloth. You can moisten the cloth with
water.
•
To clean greasy surfaces, it is recommended to apply
Isopropanol
or
Ethanol
on a piece of
lint-free cotton cloth and gently rub the surface.
5.1.2 Cleaning the trays
NEVER
clean the trays with
Xylene
, because it might harm the plastic. We recommend to apply
Isopropanol
or
Ethanol
on a piece of lint-free cotton cloth and gently rub the surface instead. In case
the tray is damaged, use either order a new one, as inserting a deteriorated tray could lead into
unsafe loading of slides.
5.1.3 Preventive maintenance
To prevent fungus growth
•
Do not leave the equipment in an unconditioned environment with a storage temperature
above 60°C (140°F) or below 0°C (32°F), which may damage the equipment.
•
Maximum operating temperature is 35°C (95°F) with a relative humidity of less than 75%.
•
Install it in a room with proper air ventilation.
•
Do not let dust or fingerprints remain on optical surfaces.
April 3, 2014 - Rev. 1
3DHISTECH Ltd.
108
(