lower measurement range.

Before using the device, check the
housing and the measurement
cables for any damage.

Do not conduct measurements in
a damp environment. Workplace,
hands, shoes and floor must be
dry.

Before opening the housing all
measurement leads are discon-
nected from the device.

2. Description, technical data

Active analog multimeter with a
slide switch for selection of the op-
erating mode and a rotary switch to
select the measurement range as
well as a scale with the mirrored
background for parallax-free read-
ings with a zero-point adjustable to
the center or the left of the scale.
By adjusting the electrical zero-point
to the center of the scale bipolar DC
voltage and current measurements
can be conducted without worrying
about polarity. The device is ex-
tremely robust in terms of load ca-
pacity and is equipped with excel-
lent overload protection due to its
two anti-parallel diodes as well as
moving coil movement which is not
sensitive to external electromagnet-
ic fields. The safety connection sock-
ets offer protection against acciden-
tal touch contact. After approx. 45
mins the battery is automatically
switched off. The power can be re-
established by turning off and on
using the slide switch for the oper-
ating mode. The robust plastic hous-
ing and the spring-connected bear-
ing jewels of the movement guar-
antee protection against damage

caused by mechanical stress.
Measurement ranges:
Voltage measurement:

DC

Meas. range Internal resistance

100 mV

10 M

300 mV

10 M

1 V

10 M

3 V

10 M

10 V

10 M

100 V

10 M

300 V

10 M

AC

Meas. range Internal resistance

3 V

1 M

10 V

1 M

30 V

1 M

100 V

1 M

300 V

1 M

Current measurement AC/DC:

Meas. range

Voltage drop

0.1 mA

55 mV

1 mA

55 mV

10 mA

55 mV

100 mA

55 mV

1 A

53 mV

3 A

51 mV

Accuracy: DC class 2; AC class 3
Effective variables and nominal op-
erating ranges:
Temperature 0 – 40° C: ± 2% / K
Frequency for all measurement
ranges:

±2.5% at 30 Hz up to
1.5 kHz ± 5% at 1.5 kHz
up to 3 kHz

Reference conditions:
Ambient temperature: +23° C ± 2 K
Frequency:  50 to 60 Hz
Waveform:   Sinusoidal

Summary of Contents for U17450

Page 1: ...d if it is used in ac cordance with the instructions Do not operate or handle this unit incorrectly or inappropriately The device may only be used by persons who are aware of the haz ards of contact for voltages over 30 V rms and can undertake the appropriate safety precautions This also includes the appearance of unforeseen voltages e g in de fective units or charged capacitors In the case of vol...

Page 2: ...ing coil movement which is not sensitive to external electromagnet ic fields The safety connection sock ets offer protection against acciden tal touch contact After approx 45 mins the battery is automatically switched off The power can be re established by turning off and on using the slide switch for the oper ating mode The robust plastic hous ing and the spring connected bear ing jewels of the m...

Page 3: ...n The needle must be lo cated in the OFF position when the multimeter is in a hori zontal position If necessary make the corresponding adjustments with the adjustment screw 5 Check the electrical zero point Set the slide switch 4 into the position Select the measurement range the needle must be set to zero point in the center of the scale otherwise correct using the rotary knob 2 Check the battery...

Page 4: ...ended 4 7 µF 630 V it is possi ble to isolate DC voltage compo nents in an amplifier output stage for example The resultant mea surement error is less than 0 2 at a measurement frequency of 50 Hz Proceed with the measurement as stated in point 3 4 Measurement of the DC voltage components is performed as de scribed under 3 3 To avoid overloading the set mea surement range must be greater than the i...

Page 5: ...ake a reading from the lower scale 3 6 3 Measuring AC current Set the slide switch 4 to the setting Use the rotary switch 1 to select the corresponding measurement range A Connect the multimeter and take a reading from the upper scale 4 Maintenance 4 1 Cleaning Only use a paintbrush or soft tow el to clean the multimeter If stat ic electrical charge builds up on the view window this can be elim in...

Reviews: