Reference conditions: 
Ambient temperature: 

+ 23° C  

Frequency: 50...60 

Hz 

Waveform: Sinusoidal 
Environmental conditions: 
Ambient temperature: 

5°C...23°C…40°C  

Storage temperature: 

-20…70°C 

Relative humidity: 

<85% with no con-
densation 

Shock tests: 

max. 147 m/s² 

Electrical safety: 
Safety stipulations: 

EN 61010-1 

Overvoltage category: 

CAT III max. 300 V; 
CAT II max. 600 V 

Degree of pollution: 

Protection class: 

II 

Protection type: 

IP20 

Overload protection: 

Fuse FF 3, 15 A / 
600 V (IEC127 6.3 
x 32 mm) 
Breaking capacity: 
1.5 kA 
Recommended type: 
SIBA: 7012540.3.15 

Electromagnetic compatibility: 
Jamming:: EN 

500081-2 

Interference immunity: 

EN 500082-2 

Power supply:: 

1 x 1.5 V Battery 
IEC LR6 

Connectors: 

4-mm safety sock-
ets 

Scale length: 

85 mm 

Pointer deflection: 

0…100°  

Operating alignment: 

horizontal 

 
Dimensions:: 98x138x35 

mm 

Weight: 

approx. 0.25 kg 

 
 

4. Operation 

4.1  Readying for use 

  Insert battery into the battery compartment. 

To do this open the appropriate section of 
the housing by pressing in the nub (7), e.g. 
using a screwdriver. Then insert the battery 
and connected it to the battery clip. Replace 
the section of housing and snap it into place. 

  Check the mechanical zero-point. The 

measuring instrument may not be connected 
to anything at this time. The needle must be 
located in the zero-point position when the 
multimeter is in a horizontal position. If nec-
essary make the corresponding adjustments 
by means of the adjustment screw (6). 

  Check the full scale deflection setting for 0 

To do this set the rotary switch (1) to “x 1 

”. Shortcircuit the connection sockets 

“COM” (5) and “A, 

” (4). Set the full scale 

deflection by turning the knob (2). 

  If full scale deflection cannot be set or the 

needle no longer stays still, the battery must 
be replaced.  

 

4.2 General instructions 

  When performing measurements always set 

the rotary switch (1) to the highest meas-
urement range. Then turn to lower ranges 
until the optimum needle deflection is ob-
tained. 

  When the multimeter is not in use, discon-

nect all measurement leads from the meter, 
reset the rotary switch (1) to the highest 
range and, if necessary, remove the battery. 

 

4.3 Voltage measurement 

4.3.1 DC voltage up to 100 mV 

 

  Set the rotary switch (1) to the measurement 

range “50 

μ

A, 100 mV”. 

  Connect the multimeter and take a reading 

from the V, A DC scale. 

 

4.3.2 DC voltage up to 600 V 

 

  Using the rotary switch (1) select the corre-

sponding measurement range “600,...,1 V 
DC”. 

  Connect the multimeter and take a reading 

from the V, A DC scale. 

 

4.3.3 AC voltage up to 600 V 

 

  Using the rotary switch (1) select the corre-

sponding measurement range “600,...,10 V 
AC”. 

  Connect the multimeter and take a reading 

from the V, A AC scale. 

Summary of Contents for AM51

Page 1: ...each of children When the multimeter is used by teenagers trainees etc a suitable person should super vise to ensure the equipment is used safely If measurements are made where there are any risks of coming into contact with elec tricity a second person is to be informed Be particularly careful when measuring volt ages in excess of 33 V AC RMS or 70 V DC For voltages which are in excess of 33 V AC...

Page 2: ... as moving coil move ment which is not sensitive to external electro magnetic fields The safety connection sockets offer protection against accidental touch contact The robust plastic housing and the spring connected bearing jewels of the movement guarantee protection against damage caused by mechanical stress 3 Technical data Measurement ranges Voltage measurement DC Measurement range Internal re...

Page 3: ...ng at this time The needle must be located in the zero point position when the multimeter is in a horizontal position If nec essary make the corresponding adjustments by means of the adjustment screw 6 Check the full scale deflection setting for 0 Ω To do this set the rotary switch 1 to x 1 Ω Shortcircuit the connection sockets COM 5 and A Ω 4 Set the full scale deflection by turning the knob 2 If...

Page 4: ...ernal voltages falsify meas ured values When resistances measurements are car ried out involving longer periods and switch over to other measurement ranges is com plete check for 0 Ω full scale deflection and adjust if necessary 4 6 Attenuation and gain measurement In communications engineering the gain or attenuation of a signal is specified in deci bels as the logarithm of the ratio between the ...

Page 5: ...se The multimeter is equipped with a safety fuse FF 3 15 A 600 V breaking capacity 1 5 kA To replace the fuse the device must be opened as described in section 4 1 Re move the fuse and replace it with a fuse of identical type Then replace the section of housing and snap it back into place se curely 6 Disposal The packaging should be disposed of at local recycling points Should you need to dispose ...