  Before opening the housing all measure-

ment leads are disconnected from the de-
vice. 

  Do not dispose of the battery in the regular 

household garbage. Follow the local regula-
tions. 

 

1.1 Symbol legend 

  Hazard, read instruction sheet 

V

 Voltage 

A

 Current 

 

Moving coil galvanometer 

  AC/DC quantities accuracy class 2,5  

 

Use in horizontal position 

 DC 

quantities 

 AC 

quantities 

CAT  Measuring category as per IEC EN 

61010-1 

 Double 

insulated casing 

CE EU 

conformity 

mark 

 Earth 

symbol

 

 

 Ground 

symbol 

 Battery symbol 

 
 

2. Description 

Handheld meter used to measure current, volt-
age and resistance as well as gain levels or 
attenuation e.g. in chains of quadripoles. 
Passive analog multimeter with a rotary switch 
for selecting the measurement ranges and a 
scale with mirrored background for parallaxfree 
readings. The device is extremely robust in 
terms of load capacity and is equipped with ex-
cellent overload protection due to its two anti-
parallel diodes as well as moving coil move-
ment, which is not sensitive to external electro-
magnetic fields. The safety connection sockets 
offer protection against accidental touch contact. 
The robust plastic housing and the spring-
connected bearing jewels of the movement 
guarantee protection against damage caused by 
mechanical stress. 

 
 
 
 
 
 

3. Technical data 

Measurement ranges: 
Voltage measurement: 

DC 

Measurement range 

Internal resistance 

100 mV 

2 k

 

1 V 

20 k

 

10 V 

200 k

 

30 V 

600 k

 

100 V 

2 M

 

300 V 

6 M

 

600 V 

12 M

 

 

AC 

Measurement range 

Internal resistance 

10 V 

66.7 k

 

30 V 

200 k

 

100 V 

667 k

 

300 V 

2 M

 

600 V 

4 M

 

 
Current measurement: 

DC 

Measurement range 

Voltage drop 

50 µA 

100 mV 

1 mA 

500 mV 

10 mA 

500 mV 

100 mA 

500 mV 

1 A 

590 mV 

 

AC 

Measurement range 

Voltage drop 

3 mA 

1.5 V 

30 mA 

1.6 V 

300 mA 

1.6 V 

3 A 

1.8 V 

 
Widerstandsmessung: 

Rotary 

switch  

Measurement range 

and scale centre 

Max. meas. 

current 

 x 1 

...35 k

...5 k

 45 

mA 

 x 10 

10 

...350 k

...50 

k

 

4.5 mA 

 x 100 

100 

...3.5 k

...500 

k

 

0.45 mA 

 
Accuracy:: Class 

2.5 

Influencing variables and nominal operating 
ranges: 
Temperature 0 – 40° C: 

± 1% / 10 K for DC 

± 2.5% / 10 K for 100 mV/50 µA DC 

 

± 1.5% / 10 K for AC 

Frequency (30 Hz...1 kHz):  ± 2.5%

Summary of Contents for AM51

Page 1: ...each of children When the multimeter is used by teenagers trainees etc a suitable person should super vise to ensure the equipment is used safely If measurements are made where there are any risks of coming into contact with elec tricity a second person is to be informed Be particularly careful when measuring volt ages in excess of 33 V AC RMS or 70 V DC For voltages which are in excess of 33 V AC...

Page 2: ... as moving coil move ment which is not sensitive to external electro magnetic fields The safety connection sockets offer protection against accidental touch contact The robust plastic housing and the spring connected bearing jewels of the movement guarantee protection against damage caused by mechanical stress 3 Technical data Measurement ranges Voltage measurement DC Measurement range Internal re...

Page 3: ...ng at this time The needle must be located in the zero point position when the multimeter is in a horizontal position If nec essary make the corresponding adjustments by means of the adjustment screw 6 Check the full scale deflection setting for 0 Ω To do this set the rotary switch 1 to x 1 Ω Shortcircuit the connection sockets COM 5 and A Ω 4 Set the full scale deflection by turning the knob 2 If...

Page 4: ...ernal voltages falsify meas ured values When resistances measurements are car ried out involving longer periods and switch over to other measurement ranges is com plete check for 0 Ω full scale deflection and adjust if necessary 4 6 Attenuation and gain measurement In communications engineering the gain or attenuation of a signal is specified in deci bels as the logarithm of the ratio between the ...

Page 5: ...se The multimeter is equipped with a safety fuse FF 3 15 A 600 V breaking capacity 1 5 kA To replace the fuse the device must be opened as described in section 4 1 Re move the fuse and replace it with a fuse of identical type Then replace the section of housing and snap it back into place se curely 6 Disposal The packaging should be disposed of at local recycling points Should you need to dispose ...